Temporary fix

?

All,

?

Based on some thoughts delivered by this thread I have found a temporary fix, but it is really only that:

?

-????????? Set SA working

-????????? Inject known power into Ant port

-????????? Set to 1dB/div and establish the error (needs the precision of being on 1dB/div)

-????????? Go to RF Analysis screen

-????????? Inject known power into main port and establish that error

-????????? Go into CONFIG screen

-????????? Set port offsets ¡®ON¡¯

-????????? Set those error offsets in the Ant port and Main port windows, to 2 decimal places

-????????? Save these settings to internal non-volatile memory as a pre-set by a useful name. I use ¡®CALFIX¡¯

-????????? As the CONFIG screen port offsets do not survive power reset, on first turning the instrument on, recall ¡®CALFIX¡¯

?

The instrument is now reading far more accurately on both Ant input dBm and the Power Meter, and these can all be done from the front panel.

?

There remains another offset on SA accuracy from the Main RF port that can¡¯t be removed without also affecting the Power Meter.

?

But as I said, it¡¯s not a particularly desirable or elegant solution, though I hope this is helpful to some

?

Rgds

?

Colin G4CWH/ZF2CA/M3E