Re: HP 8920A spectrum analyser amplitude calibration
Temporary fix
?
All,
?
Based on some thoughts delivered by this thread I have found a temporary fix, but it is really only that:
?
-?????????
Set SA working
-?????????
Inject known power into Ant port
-?????????
Set to 1dB/div and establish the error (needs the precision of being on 1dB/div)
-?????????
Go to RF Analysis screen
-?????????
Inject known power into main port and establish that error
-?????????
Go into CONFIG screen
-?????????
Set port offsets ¡®ON¡¯
-?????????
Set those error offsets in the Ant port and Main port windows, to 2 decimal places
-?????????
Save these settings to internal non-volatile memory as a pre-set by a useful name. I use ¡®CALFIX¡¯
-?????????
As the CONFIG screen port offsets do not survive power reset, on first turning the instrument on, recall ¡®CALFIX¡¯
?
The instrument is now reading far more accurately on both Ant input dBm and the Power Meter, and these can all be done from the front panel.
?
There remains another offset on SA accuracy from the Main RF port that can¡¯t be removed without also affecting the Power Meter.
?
But as I said, it¡¯s not a particularly desirable or elegant solution, though I hope this is helpful to some
?
Rgds
?
Colin G4CWH/ZF2CA/M3E
|