¿ªÔÆÌåÓýHi Mark ? From the results given it could be that as stated earlier it requires realigning The master yig is multiplied several times to give the full band? ? so during alignment the switching between the bands requires setting so they do not overlap or leave gaps and it is this measurement you require to monitor the sweep with a scalar analyser ? ? but in my early days I used a diode detector into a scope with the sweeper supplying the X axes this is best dun with an attenuate and not levelled as the bands stand out easier ? during the alignment you might have to change resisters? this will require the use of a decade resistor box sorry its been over 25 years since I did one ? Paul From:
[email protected]
[mailto:[email protected]]
On Behalf Of Mark Bielman ? Unit just arrived.
Initial testing does not agree with original complaint: "Output at 2Ghz is -10db below what it should be. Won't
produce a signal much past 10Ghz and at 10Ghz it is -30db below the RF output
level setting. No output past this frequency" No virus found in this message. |