Re: JOHN E FAST


 

On my 720 and 717, drifting appears to be contamination on the high ohm resistors next to the triode under the cover. I looked at the procedure to clearn those and I feel I better leave that alone. When you add desiccant, and it behaves a little better, that tells me there is moisture reacting with stuff on those resistors causing leakage currents.

Team: Does KI4U also fix units that have issues or only provide calibration?

-Stephen N.

On Mon, Jan 30, 2023 at 9:02 PM Clint <clintb@...> wrote:
Yes.? I ruled out the second and third things shown below leaving possible moisture contamination.? I tried desiccant but it didn't completely fix the problem.? I would like to have more confidence that it will pass calibration before sending it in.? I have a couple of meters that pass the circuit check and have a steady needle on all ranges, but that doesn't guarantee they can be calibrated.? Maybe I'm mistaken but I doubt a meter with a drifting or erratic needle will pass.?



Looking at the John E Fast cap it appears to have a porous ceramic casing, so I was curious if anyone was aware of a problem with those capacitors before swapping them out.



Clint



--
Stephen Nelson