Yes it has to run for a good long time to get your stable reading, when first started pumping it reads very low, orders of magnitude.
NICK- ask this question of your contact - what OTHER isotopes can it be calibrated to? Today I tried changing the Regions of interested or shutting them off, but it wouldn't seem to let that happen. I suspect we can block the Uranium generated radon from alarming and just have the Thorium radon set it off?
I think I'll let it smell Thoron gas for a while and watch the spectrum.
Geo
toggle quoted message
Show quoted text
----- Original Message -----
From: Dude <dfemer@...>
To:
[email protected]Sent: Mon, 30 May 2022 15:48:04 -0400 (EDT)
Subject: Re: Alpha and Beta particle nuclear decay studies. Alpha and Beta particle Spec
Geo,
Yes 1.8 pCi/l you just
have to multiply the screen reading by 1000 to get Liters.
How long was that run when
you took the picture, 53 min or so? and what was the flow rate ?0.87 l/m?
?That low radon count must be from a limestone bedrock or you opened a
window.
Dud
?
?
?
doesn't that work out to
1.8pCi/L?
That's about right for
here, I think. Normal background Gamma is low here in particular because behind
the house is no ground, just air.
----- Original Message
-----
From: Dude <dfemer@...>
To: [email protected]
Sent: Mon, 30 May 2022 13:32:55 -0400 (EDT)
Subject: Re: Alpha and Beta particle nuclear decay studies. Alpha and Beta
particle Spec
margin: 0.0in;
font-size: 12.0pt;
font-family: "Times New Roman" , serif;
}
a:link, span.MsoHyperlink {
color: blue;
text-decoration: underline;
}
a:visited, span.MsoHyperlinkFollowed {
color: purple;
text-decoration: underline;
}
p {
margin-right: 0.0in;
margin-left: 0.0in;
font-size: 12.0pt;
font-family: "Times New Roman" , serif;
}
p.MsoNoSpacing, li.MsoNoSpacing, div.MsoNoSpacing {
margin: 0.0in;
font-size: 16.0pt;
font-family: Arial , sans-serif;
color: black;
}
span.EmailStyle19 {
font-family: Arial , sans-serif;
color: black;
top: 0.0pt;
letter-spacing: 0.0pt;
font-weight: normal;
font-style: normal;
}
*.MsoChpDefault {
font-size: 10.0pt;
}
div.Section1 {
page: Section1;
}
/*]]>*/
Geo,
You may have to
recalibrate the touch screen . Mine was very hard to get the monitor tab to
come up. There is a procedure in the manual that will allow you to fix the
touch
screen positions. I found using a soft pointed touch screen pen worked better
than fingers on that top bar.. Be careful not to scratch the screen with a hard
pointy pen.
?
How long was that run when
you took the picture 53 min or so? and what was the flow rate ?0.87 l/m?
?That
low radon count must be from a limestone bedrock or you opened a window.
Dud
?
?
It took more time to
figure out you can't punch those buttons easily with FF (Fat Fingers) Tried
some tools and found something that worked AOK. Following your instructions was
easy (sort of, regular easy, like training a cat easy, not easy-peasy).
Results are in line with
the other instruments now. Those have been taken in other rooms though, but
very close.
Will set them all up in
same place and see for real.