Re: #about_wiki
#about_wiki
Hello,
Here: I can't understand phrase: "One marker is active and marked with an inverted number." I have never seen inverted numbers on a marker. In what modes does this happen? Or does it mean something like "The active marker is marked with a number in an inverted triangle."? With best regards, |
Re: #about_wiki
#about_wiki
Thanks. Good remark. It's indeed maximum 290. But there is an option to select less sweep points
-- For more info on the tinySA go to https://tinysa.org/wiki/ |
Re: #about_wiki
#about_wiki
Better: " As the number of points displayed is 290, when the RBW is too small to cover...."
|
#about_wiki
#about_wiki
Hello,
In the I read the phrase: The number of points scanned is minimum 290 but when the is too small to cover the selected frequency range with 290 points additional measurements are done and accumulated into the 290 points. Is the highlighted word correct? According to the specifications of the tinySA3, the word "maximum" is more appropriate. With best regards, |
Re: Using TinySA with VMA TinySA
On Thu, Jan 26, 2023 at 01:40 AM, Vitor Martins Augusto wrote:
Perhaps Eric could include a START/STOP command, so that external software could switch off the screen of the TinySA and stop it doing its own sweeps while being used remotely. The external software would then either reactivate the screen with START or the user would have to switch teh TinySA off and back on.You can use the pause and resume command. When in pause mode a big green PAUSED text is shown in the status window ? -- For more info on the tinySA go to https://tinysa.org/wiki/ |
Re: Using TinySA with VMA TinySA
Also, the TinySA will stop his own sweep, when the raw command is sent through the USB/RS232 protocol. Once the raw command finishes its sweep, the TinySA resumes what it was doing. But since my software constantly sends raw commands, the TinySA ends up showing outdated sweeps, when it does not get time to do its own sweep. Also, the use of the raw command will not change the settings for TinySA's own sweep, so you end up seeing something completely weird.
Perhaps Eric could include a START/STOP command, so that external software could switch off the screen of the TinySA and stop it doing its own sweeps while being used remotely. The external software would then either reactivate the screen with START or the user would have to switch teh TinySA off and back on. |
Re: Using TinySA with VMA TinySA
Hello,
Please e-mail me directly with problems regarding my software - this group should be use for the TinySA devices only, as per request from Eric, which I respect. The mentioned problem results from having "," configured as the decimal symbol. Visual Studio is IMHO not coherent regarding the use of regional settings, some functions respect it, some don't. The fix consists in using "." as the decimal symbol. I will try to fix this in future versions. Regards, Vitor |
Re: #feature_request
#feature_request
LV Zeners make wonderful noise sources. Dave - W?LEV On Wed, Jan 25, 2023 at 9:50 PM John Cunliffe W7ZQ <n2nep@...> wrote: But then, noise sources are easy to build and used externally or bought cheap on the internet. Why bogging down the TinySA with it? -- Dave - W?LEV |
Re: #feature_request
#feature_request
But then, noise sources are easy to build and used externally or bought cheap on the internet. Why bogging down the TinySA with it?
|