The upper part of the signal goes up when I activate the lna
The lna is supposed to be activated to see more of the lower part of the signal, but in my case, apart from seeing more of the lower part of the signal, the upper part goes up 3 or 4 db, so I don't know what the true value is. The signals I'm measuring have values ??between -70 and -60 dbm. What am I doing wrong? |
Re: Self Test failures from New TinySA Ultra --- Amazon purchase of ZeenKo unit from SDRGEEK Store
On Wed, Nov 20, 2024 at 04:52 PM, Alex P wrote:
Plugging it into a windows 10 computer should automatically load the serial over USB drivers needed to communicate with the tinySA. To control the tinySA from the PC you have to download some SW, see:
Download the PC SW on the XP machine and see if it starts.?
--
Designer of the tinySA For more info go to |
Re: Self Test failures from New TinySA Ultra --- Amazon purchase of ZeenKo unit from SDRGEEK Store
I think you need to read this On Thu, Nov 21, 2024 at 12:52?AM Alex P via <winston376=[email protected]> wrote:
|
Re: Self Test failures from New TinySA Ultra --- Amazon purchase of ZeenKo unit from SDRGEEK Store
Erik:
?
Thanks....I'll be returning it in the coming days.
?
Although I have indeed read the wiki? and looked at some videos,? I do need some additional guidance and info....I would want to interface it with my? Windows 10? and I'm unclear as to whether just plugging the genuine? Ultra? instrument will present a GUI upon insertion into the USB port????
?
Also, does it function on a Windows XP....I've got several of these with decades of technical software, and programs that were/are no longer supported and only work on Windows.
?
Thanks
?
Alex |
Re: Wrong implementation of the quasi peak measurement
#bug_report
For an unmodulated signal, the QP and peak values should be identical.? And, yes, the "smoothing" is done in the time domain, not frequency domain.? I have always viewed QP as a post detection averaging or "filter". In the two screen grabs, the peak levels are pretty much the same, but the exponential tail should not be present.? Yes, there is something not quite right with the coded UP routine. Dave - W?LEV On Wed, Nov 20, 2024 at 8:27?PM gajdacsa via <gajdacsa=[email protected]> wrote:
--
Dave - W?LEV |
Re: v1.4-185 Changelog SA ultra
please have a look at this message:?
/g/tinysa/topic/new_fw_release_load_trace/109408685
message: /g/tinysa/message/18473 ?= version V1.4184 /g/tinysa/message/18486 : V1.4185?
? ?updates: /g/tinysa/message/18497 ? ?and ?/g/tinysa/message/18506 |
Wrong implementation of the quasi peak measurement
#bug_report
Hi,
?
I took 2 screen dumps to demonstrate that the quasi peak measurement method does some sort of frequency domain smoothing as of the current implementation, which is clearly not what it should be doing. The 1st screenshot was made using the AVE16 calc method, the 2nd is QP. They should be nearly identical since?I measured a stationary signal.
?
The quasi peak?smoothing must be done in the time domain, so each frequency bin of interest must be measured separately for a certain amount of time and the?consequently taken samples need to be fed to a quasi peak detector with an attack and release time specified by?CISPR.
?
Best regards,
Andras
?
?
|
Re: Self Test failures from New TinySA Ultra --- Amazon purchase of ZeenKo unit from SDRGEEK Store
Ignore the post about the 30 kHz signal, it should be 30 MHz and its automatic when doing a selftest.
?
Your tinySA (clone?) is very bad quality, strongly suggest you return it and buy from authorized seller such as R&L instead
--
Designer of the tinySA For more info go to |
Re: Self Test failures from New TinySA Ultra --- Amazon purchase of ZeenKo unit from SDRGEEK Store
Cable is connected correctly.... CAL to REF.
?
Self test freezes at test #1...."Above Critical".
?
Repetitive screen touching advances the 'self test' routine.
?
The following tests 'Passed':? 6, 8, 9, 12, 13.
?
All others either 'Signal Level Fail': (3, 4, 7, 10, 11, 14) or? 'Above critical': (1, 2, 5,).
?
Where do I find the 30 KHz Cal signal that you are referring to?? What's the path to it?
?
?
Additional anomalies:
?
Brightness adjustment default is 70%.?? Changing the brightness with the jog wheel freezes the instrument.
?
?
Additional question:
?
The various online info infers that I can use a Windows (10) OS to interface with the instrument as soon as I insert the instrument to my USB port and the kernel/driver is supposed to be self-executable.?
?
What is supposed to happen upon insertion of the device to my USB port?
?
?
Thanks.
?
Alex
?
? |
Re: Choosing an attenuator
dBm is shorthand for dBmW, with dBm as typical unit for Rx ports on measuring instruments.
dB is just a scalar, for attenuators we still need the power handling spec. 30dB 10w, 20dB 5w, etc.
I guess along the way the Rf folks stole the minus sign and shoved it into the word "attenuate". In technical terms, the actual electrical result is? -dB.
?
-dB attenuate, "attenuated 30dB"
+dB amplify, "amplified 30dB"
?
Such oddity in words. ;)
?
I am curious about OP's transmitter. Is the Tx port 50 or 75 ohm? |
Locked
Re: Possible design issue with TinySA Ultra.
On 15/11/2024 15:20, The Greene Family via groups.io wrote:
"the improved tinySA ULTRA *plus*"There is some unofficial information published like: -- Arkadiusz Mi?kiewicz, arekm / ( maven.pl | pld-linux.org ) |
Re: Self Test failures from New TinySA Ultra --- Amazon purchase of ZeenKo unit from SDRGEEK Store
Hi again FYI;?
Are the only stores on Amazon that are on the approved list. Clyde KC7BJE? On Tue, Nov 19, 2024, 10:07?PM Alex P via <winston376=[email protected]> wrote:
|
Re: Self Test failures from New TinySA Ultra --- Amazon purchase of ZeenKo unit from SDRGEEK Store
SDRGREEK STORE is not on the approved WiKi list of dealers it may be possible that you may, have a knock-off or clone.
Which selftest failed? Did you preform the signal level setup? Did you connect port 1 and 2 correctly? Did you turn on the calibration signal to 30khz?
You need to give more info than just the selftest failed.
Clyde KC7BJE
?
? |
Self Test failures from New TinySA Ultra --- Amazon purchase of ZeenKo unit from SDRGEEK Store
Just received the TinySA Ultra and it fails the self test.
?
Didn't see all of the comments about clones until I tried to set the unit up,? but need to know if any other users have any experience from this ZeenKo unit from this Amazon seller,? SDRGEEK store?
?
Thanks
?
Alex |