Self Test Fail
Thanks for allowing me to join. I just received my Tiny SA this week and performed the self test diagnostics. Unfortunately there were errors present during this. I'd like to know the issue that I'm seeing and how to correct the errors. Please see attached screen shot and advise.
Thanks Joe |
Re: New tinySA failing Self-test after FW upgrade
On Wed, Jan 20, 2021 at 12:25 PM, Nicholas R wrote:
For interest sake, I ran the new test...Nicholas, What type of sources have used in combination with the tinySA? Any sources that generate possibly more than +10dB? ? -- ------------------------------------------
For more info on the tinySA go to https://tinysa.org/wiki/ |
Re: What to do when a selftest fails
On Thu, Jan 21, 2021 at 08:47 PM, Tom D'Orsay wrote:
Ran 30db test twice and it failed badly 4 to 5 db variation.Tom? What type of sources have used in combination with the tinySA? An sources that generate possibly more than +10dB? ? -- ------------------------------------------
For more info on the tinySA go to https://tinysa.org/wiki/ |
Re: What to do when a selftest fails
? ? ? ? ? ? ?Erik:??
So I upgraded to 1.1-77. Had to manually install drivers from win64 drivers folder.? Ran 30db test twice and it failed badly 4 to 5 db variation. I did a recalibrate after firmware update and had saved reference info and checking found it still the same. Looks like I ned to return this to R&L for exchange. I notified them a few days ago that I was testing further and would advise. ? ? ? ? Tom ? ? ? ? KA1GXR |
Re: NEw FW: Linearity test enabled
First run unit does nicely in this test
73 -Jim NU0C On Tue, 19 Jan 2021 23:53:00 -0800 "Erik Kaashoek" <erik@...> wrote: New FW released |
Re: New tinySA failing Self-test after FW upgrade
On Tue, Jan 19, 2021 at 03:33 PM, hwalker wrote:
I noticed that 0.5 dB steps can be entered from the on-screen keypad but only only 1 dB steps can be programmed using the 'attenuate' command (i.e. attenuate 15.5 displays Atten: 15dB on the tinySA status display.? Is that an intentional limitationHerb. I was not sure 0.5dB steps where needed and it is working by accident in the tinySA ? -- ------------------------------------------
For more info on the tinySA go to https://tinysa.org/wiki/ |
Re: What to do when a selftest fails
开云体育Eric:The test on my unit seems to start at 29dB of attenuation. charlie On 1/20/2021 2:42 AM, Erik Kaashoek
wrote:
It would help us if you could install the a new FW. V1.1-77 and run an additional test |