Hi,
I've just updated firmware to v1.4-159 on both my tinySA and my Ultra and both failed self test after that.
TinySA reported "Test 7: StopBand Critical"

Ultra failed both Test 10 and 14, both reporting "Signal Level Fail"

Both units were tested using the longer ALSR-100 cable provided with the Ultra version.
I have also tried connecting both together, outputting a 300MHz, -5dBm signal from the tinySA and reading it at around -6,6dBm on the Ultra (wich seems about correct to me), when I tried doing this with the shorter cables (the one that came with the tinySA) the Ultra read -9.0dBm
Both units should be genuine as they were bought from Eleshop.eu following link from the wiki
tinySA was bought in May 2022, HW version: V0.3.1_EL
Ultra was bought in Dec 2022, HW version: V0.4.5.1
Have I broken something or missed some step?
Thanks!
--
//LMA//
RF Coordinator for live shows