Re: -H not usable #problem
All, Thanks for a lot of good thoughts and info. I have done all except I did not add any washers: there was no difference with the case off. After all the tries, it seems to be working even though
By Paul W8SBH · #13903 ·
VNA Saver and Win 8.1 Connection Problem
Hello, Perhaps someone on here could point me in the right direction. I have been given an old HP laptop running 8.1. It would be perfect to use outside with my nanoVNA. I have two Win 10 desktop
By Paul Sayer · #13902 ·
Re: NooElec 1:9 balun #test-jig
Hello Roger, My answers are in yout text I postet it already in /g/nanovna-users/message/13893 See: NooElelc_Balun_9-1_S21_Loss.png Yes, I have also a No-Name version of the Balun.
By Rudi · #13901 ·
Re: NooElec 1:9 balun #test-jig
Rudi, Glad to hear you got the results you expected A couple of questions... 1. Your concern was that S11 was different when you hooked up the NanoVNA instead of the 50 ohm load. Could you please post
By Roger Need · #13900 ·
Re: -H not usable #problem
I find using a soft stylus made a capacitance touch display or the eraser end of a mechanical pencil works better than my finger. See attachment. Don't use the hard pointed end of the stylus which is
By Mike N2MS · #13899 ·
Re: -H not usable #problem
That is typical of screen cal. Use the jog sw to select the screen cal function, cal and test but remember to save the cal through the menu.
By Larry Rothman · #13898 ·
Re: -H not usable #problem
I tried to cal the touch screen. The menu bar is erratic. Touch it and it usually disappears. I wonder if this is a screen cal issue or if it is something else (softwre?). wrote: -- Paul W8SBH ??
By Paul W8SBH · #13897 ·
Re: Test fixture question
Thanks for the compliment and the advice. Attenuators are a design feature, component values are on he schematic for 3, 6, and 15 dB. I discovered a similar comment while researching various ideas in
By Bill Buoy - N5BIA · #13896 ·
Re: Nanovna-H4 Output Shuts off after Cal #calibration
Hi Mark, at first I didn't understand your question, then it started to dawn, maybe you have the touch screen doing this thing too. The touch screen is tensioned or pressed somewhere, or the cover
By Gyula Molnar · #13895 ·
Re: -H not usable #problem
My suggestion is to remove the screen-side cover plate and see if the error phenomenon still exists. The touch screen is particularly sensitive anywhere there is tension or pressure on the surface.
By Gyula Molnar · #13894 ·
Re: NooElec 1:9 balun #test-jig
Hello David and Roger, This was the solution. Thank you very much for the hint. I have the schematic of the balun, but I did not consider the R1 jumper. II cut the R1 jumper on the lower side of the
By Rudi · #13893 ·
Re: NooElec 1:9 balun #test-jig
David had a good suggestion on cutting the trace on that NanoVNA balun. I have done that on the ones I own. Here is a circuit drawing of the balun and a photo showing how to cut the track. Roger
By Roger Need · #13892 ·
Re: NooElec 1:9 balun #test-jig
The high impedance output of this balun is center tapped and the center tap is grounded back to the input SMA (unless you cut the small trace on the back of the PCB). When you connect one side of the
By Dave W6OQ · #13891 ·
Re: NooElec 1:9 balun #test-jig
Calibrate at the end of the 50cm of cable! USe a small SMT resistor 49.9 ohm 1% works well enough or you can get exact 50 ohm in 1% SMT. Short and open are easy. Did this kind of stuff for measuring
By aparent1/kb1gmx <kb1gmx@...> · #13890 ·
Re: -H not usable #problem
Have similar issues with my new H4. The touch screen interface is inconsistent. Sometimes pressing a menu action works with the correct option highlighted. Other times not either with the wrong button
By Simon Eatough <simon.eatough@...> · #13889 ·
Re: -H not usable #problem
Have similar issues with my new H4. The touch screen interface is inconsistent. Sometimes pressing a menu action works with the correct option highlighted. Other times not. I have calibrated the LCD.
By Simon Eatough <simon.eatough@...> · #13888 ·
Re: -H not usable #problem
If the touch screen still fails to work after recalibration, the touch screen may be damaged and you can contact your dealer for replacement.
By Hugen · #13887 ·
Re: -H not usable #problem
Possible Need run touch screen calibrate Push leveler button Push right while not select CONFIG menu Push leveler button Push left or right to select TOUCH CAL Push leveler button On screen use stylus
By DiSlord · #13886 ·
Re: NooElec 1:9 balun #test-jig
Rudi, Here is a suggestion - Try the test again by putting the 50 ohm load at the end of the cable instead of connecting it to CH1. Now you just have the coaxial cable connected with the same load on
By Roger Need · #13885 ·
-H not usable #problem
I have the -H model, HW ver. 3.4, SN 20011259 build Dec 26, 2019. I have had this for about 2 weeks and it had been working nicely. Purchased from GigaParts. 1. The unit turns on and the
By Paul W8SBH · #13884 ·