try to do your sweeps with some averaging (means more points/values per measure point are takern and interpolated) ... that make the curves more even
you even can do the calib with high avg .. and measure then with lower avg (on none at all)
are your connectors tight?? what cable do you use?? are those values jumping around if you move cables a bit (wiggle on them)???
that you have those waves at the lower end and straight lines on the rest of the scan is not normal
dg9bfc sigi
Am 30.08.2022 um 19:11 schrieb TE Dukes:
toggle quoted message
Show quoted text
Thanks!!
I watched the video and did the calibration along with the guy. My menus are a little different than his. I don't have the choice of 'CORRECTION'. I think the equivalent on mine is 'Apply".
Anyway, the points are where they are supposed to be BUT I have green lines coming out of each point jumping around. Pretty sure its not supposed to be doing that.
Any ideas?
Thanks