开云体育

ctrl + shift + ? for shortcuts
© 2025 Groups.io

Re: Trinitite

 
Edited

Refer message 766 and 772

Analyzing Trinitite with emphasis on the Uranium and Neptunium Ly1 peaks.

Preparation; First I did the analysis of Uranium K and L lines mentioned in message #766. Then I did same for an Am-241 source. Next the two were overlaid and notations made to show the subtle but measurable shift in the
La and Lb lines and the more noticeable shift in the Ly1 lines.(y= gamma, third letter in Greek).

Am_vs_Np-L-Lines-small.jpg


So all that is needed to finish the experiment is a good scan of Trinitite on the Si-PIN with the same settings as the?
Am-241, and those two graphs overlaid. This one happens to be some Red Trinitite, as the big green one is in another chamber being tested on a long 24 hour run.

Red-Trinitite-VS-Am-241-20FEB-2020-U-VS-NP-Ly1-Notes-small.jpg



Why does it matter if the lines are U or Np? It is just one step in testing Trinitite for Plutonium.?
we already know Am-241 is a decay product of Pu-241, and there are only 3 easy ways to get Uranium X-Rays-
1:Decay of U-238, to the stage where U-234 is created
2: Deliberate XRFing Uranium, or unintentionally XRFing it by a random energy source.
3: Decay of Plutonium or one of its higher progeny to Uranium.

Remember, natural decay induced X-Rays come from the daughter at the moment of creation, not from the parent.

How many other ways can we test for the presence of Plutonium? Any species will do, as long as its half life is long enough for there to be some left.

Hint: one way to at least make progress is Alpha Spectroscopy. My plans are to marry a silicon Alpha detector to an Amptek DP5 processor and hope for results. An external preamp is needed, I have those in NIM, discreet, removed from surplus equipment and in this ingenious device made by a company called The Nucleus.

Alpha-Beta-Spectrometer-5030-open-vacuum-chamber-small.jpg

Equipment manufacturer by Spectrum Techniques today is so similar to other models made by The Nucleus that I suspect a link there.





Geo>K0FF




Re: Testing for Cd in Zn Ore.

 

开云体育

Geo,

Sphalerite (ZnS) has an affinity to pick up a lot of different impurities. Here in southern Az Cd is common as one of those impurities. Try running some Sphalerites’ and see what you get.

I also had some Cd in Galena (PbS) at 89 ppm with the Ag running 0.7% (234 oz/ton).?

Dud

From: [email protected] [mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Tuesday, February 18, 2020 7:37 AM
To: [email protected]
Subject: [XRF] Testing for Cd in Zn Ore.

?

Thanks to member Steve D. in ALBQ NM for mentioning that Zn ores are also a common source of?
Cd.

Testing several species of ore here with the new emphasis on Cd has proved problematic. First the X-Ray hump from the tube is most intense right where Cd K line XRF is located, making it harder to pick out a weak peak there. Strong K peaks- no problem, as shown in one of the pictures to follow.

So I tried looking for L peaks from Cd in? ore.?

First steps are to use pure Cd metal to establish calibration points exactly for 3.2, 3.13, 3.31, 3.52, and 3.71kev.

Then another XRF test of the empty chamber, with no samples inside, for the same time and exposure data.

In all tests for this series, the 3 Al beam filters remain but the final collimator is removed to give a wide beam on the mineral tests to follow. The detector is an Amptek Si-PIN into the Amptek DP5-PC5 electronics stack, then USB to a notebook computer for control and storage.

First the pure metal "stamp"r, X-Ray source set to 37kVp-25uA

CD-Pure-Metal-Stamp-Ka-La-xrf-test-37kVp-25uA-307s-Notes1-small.jpg


In the above, Ka, Kb are in perfect alignment and something is barely detected around 3 keV This rendition (LINEAR) shows how well the K lines are excited in the lighter elements, compared to their L lines.

Geo


Re: Si pin Detectors

 

No connection to the eBay seller- but I know the sale for the sensor, sensor cable, and DP5-PC5 PCBs comes in a factory sealed package and the power supply is correctly programmed at the factory for that unique sensor.

Buying the boards then the sensor separately leaves the user to correctly program the hardware to the sensor being used.

Another note- Si-PIN marked DP-5s won't operate Silicon Drift Detector sensors

and IMPORTANT- don't apply?operating voltage to ANY of these kind of sensors without having the sensor mounted on a HEATSINK.


Geo


Re: Si pin Detectors

 

Taray,

Just to be clear, I have modified Theremino to import and display the .mca files that are collected by DppMCA, the Amptek MCA program designed to work with their hardware.? Theremino does not collect data directly from the DP5-PC5 PCB stack.

As George mentioned, you will need more than just this detector for a working system.? It looks like about $1200 for the components, which then need to be integrated together.? George knows more about that since I bought my working system from him.

Charles


On Thu, Feb 20, 2020 at 6:58 AM <GEOelectronics@...> wrote:
Welcome Taray, sorry for the delay in your post showing up, Groups.IO makes the moderator personally approve the first message from every group member. From now on your posts will show up immediately.

Talk to Charles Young, a member here who has adapted the Amptek to the
Theremino? program and made improvements to that program.

I can comment on the hardware- how I did it- you will need the sensor, and the DP5-PC5 PCB stack from eBay, then a 5V power supply (2 amps) and a heat sink direct from Amptek. These cost about $200, and could possibly be home made, again talk to Charles before buying on eBay.

George Dowell
Geo>K0FF








----- Original Message -----
From: taray singh via Groups.Io <sukhjez=[email protected]>
To: [email protected]
Sent: Wed, 19 Feb 2020 19:30:33 -0500 (EST)
Subject: [XRF] Si pin Detectors

Hi guys

I like to get one of those Si pin detectors

There is an eBay??seller selling..

?

Si-PIN (FSJ32MD-G3SP)?detector element with PA230? Preamplifier, 25mm^2, 1 mil Be window for??about Us500?

?

It seems??to be needing connections to a Amptek MCA with preamp

Also seen Thermimo pics of usage??in our forum.

How do I??connect a??Si pin??sensor to get on Thermimo??for a complete affordable setup?It appears to be a cheaper option

Taray





Re: Si pin Detectors

 

Welcome Taray, sorry for the delay in your post showing up, Groups.IO makes the moderator personally approve the first message from every group member. From now on your posts will show up immediately.

Talk to Charles Young, a member here who has adapted the Amptek to the
Theremino? program and made improvements to that program.

I can comment on the hardware- how I did it- you will need the sensor, and the DP5-PC5 PCB stack from eBay, then a 5V power supply (2 amps) and a heat sink direct from Amptek. These cost about $200, and could possibly be home made, again talk to Charles before buying on eBay.

George Dowell
Geo>K0FF








----- Original Message -----
From: taray singh via Groups.Io <sukhjez@...>
To: [email protected]
Sent: Wed, 19 Feb 2020 19:30:33 -0500 (EST)
Subject: [XRF] Si pin Detectors

Hi guys

I like to get one of those Si pin detectors

There is an eBay??seller selling..

?

Si-PIN (FSJ32MD-G3SP)?detector element with PA230? Preamplifier, 25mm^2, 1 mil Be window for??about Us500?

?

It seems??to be needing connections to a Amptek MCA with preamp

Also seen Thermimo pics of usage??in our forum.

How do I??connect a??Si pin??sensor to get on Thermimo??for a complete affordable setup?It appears to be a cheaper option

Taray





Si pin Detectors

taray singh
 

Hi guys

I like to get one of those Si pin detectors

There is an eBay??seller selling..

?

Si-PIN (FSJ32MD-G3SP)?detector element with PA230? Preamplifier, 25mm^2, 1 mil Be window for??about Us500?

?

It seems??to be needing connections to a Amptek MCA with preamp

Also seen Thermimo pics of usage??in our forum.

How do I??connect a??Si pin??sensor to get on Thermimo??for a complete affordable setup?It appears to be a cheaper option

Taray


Re: Trinitite

 

开云体育

Geo,

That dead time is pretty high. Its best to Set the tube ?current to keep the dead time around 10% or less than 35K cps. First shot should use full voltage then set the Kv’s ?to target the absorption edge of element range of interest? say about 1.5 the times the edge. Use filters for that edge to target the element of interest. ?Good luck sorting out Ba from decay, XRF and that in the soil.

Dud

?

?

From: [email protected] [mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Tuesday, February 18, 2020 1:28 PM
To: [email protected]
Subject: Re: [XRF] Trinitite

?

Now my largest green-glass Trinitite sample tests.

Trying to determine the "best" settings to detect good XRF on the Si-PIN sensor:
This test is?XRF by X-Ray tube at different kVp levels and current levels.Other tests will use AmX8 exciter and anything else I can think of.

All the settings agree, there is considerable barium element present. Some is from Cs-137 beta decay (Ba-137m) and is responsible for the familiar 662 keV peak we all use for calibration. Some could be old fission products and some could have been in the soil before Trinity.

Ba_in_Trinitite-note.png

A few different X-Ray setting results .mca attached. There should be enough data here to really determine the sweet spot for settings, taking into account peak broadenings, deadtime, and runtime.

Geo>K0FF


Re: Trinitite

 

That makes sense too.

Geo

----- Original Message -----
From: WILLIAM S Dubyk <sdubyk@...>
To: [email protected]
Sent: Tue, 18 Feb 2020 20:58:49 -0500 (EST)
Subject: Re: [XRF] Trinitite


Some of the barium may be from Baratol, a barium nitrate/TNT mix that what used for the explosive lens for the device. Low velocity stuff.


Steve





From: [email protected] <[email protected]> on behalf of GEOelectronics@... <GEOelectronics@...>
Sent: Tuesday, February 18, 2020 2:28 PM
To: [email protected] <[email protected]>
Subject: Re: [XRF] Trinitite
?
Now my largest green-glass Trinitite sample tests.


Trying to determine the "best" settings to detect good XRF on the Si-PIN sensor:

This test is?XRF
by X-Ray tube at different kVp levels and current levels.Other tests will use AmX8 exciter and anything else I can think of.

All the settings agree, there is considerable barium element present. Some is from Cs-137 beta decay (Ba-137m) and is responsible for the familiar 662 keV peak we all use for calibration. Some could be old fission products and some could have been in
the soil before Trinity.




A few different X-Ray setting results .mca attached. There should be enough data here to really determine the sweet spot for settings, taking into account peak broadenings, deadtime, and runtime.


Geo>K0FF





Re: Trinitite

 

开云体育

Some of the barium may be from Baratol, a barium nitrate/TNT mix that what used for the explosive lens for the device. Low velocity stuff.

Steve




From: [email protected] <[email protected]> on behalf of GEOelectronics@... <GEOelectronics@...>
Sent: Tuesday, February 18, 2020 2:28 PM
To: [email protected] <[email protected]>
Subject: Re: [XRF] Trinitite
?
Now my largest green-glass Trinitite sample tests.

Trying to determine the "best" settings to detect good XRF on the Si-PIN sensor:
This test is?XRF by X-Ray tube at different kVp levels and current levels.Other tests will use AmX8 exciter and anything else I can think of.

All the settings agree, there is considerable barium element present. Some is from Cs-137 beta decay (Ba-137m) and is responsible for the familiar 662 keV peak we all use for calibration. Some could be old fission products and some could have been in the soil before Trinity.

Ba_in_Trinitite-note.png

A few different X-Ray setting results .mca attached. There should be enough data here to really determine the sweet spot for settings, taking into account peak broadenings, deadtime, and runtime.

Geo>K0FF


Re: Trinitite

 

Now my largest green-glass Trinitite sample tests.

Trying to determine the "best" settings to detect good XRF on the Si-PIN sensor:
This test is?XRF by X-Ray tube at different kVp levels and current levels.Other tests will use AmX8 exciter and anything else I can think of.

All the settings agree, there is considerable barium element present. Some is from Cs-137 beta decay (Ba-137m) and is responsible for the familiar 662 keV peak we all use for calibration. Some could be old fission products and some could have been in the soil before Trinity.

Ba_in_Trinitite-note.png

A few different X-Ray setting results .mca attached. There should be enough data here to really determine the sweet spot for settings, taking into account peak broadenings, deadtime, and runtime.

Geo>K0FF


Re: Trinitite

 

more pics of the metallic Trinitite

This is the first CdTe in its chamber. The "lid" is a lead (Pb) Half-Brick that slides over the opening.

CdTe- Gamma-Ray-Spectrum-Analysis-Both-Blob-and-Triangle-small.jpg


This is a different CdTe, same as the first one, it has its own chamber too, but much smaller, only the sample and sensor portion of the probe goes into it. Usually one is set to run 0-200 keV the other one to 0-400 keV (max range). Both run at the same time, independently.

The dots on the sample is a paint marker to identify the edge or surface being tested. Thanks to Al Shinn for that tip.

CdTe-Triangle-TrinititeGamma-Spec-One-Surface-at-a-time-small.jpg

Next post will be some XRF of the green glassy type of Trinitite- has anyone bothered to analyze the non radioactive elements present?


Geo>K0FF


All natural- U-235 Gamma Spectrum

 


U-235-All-Natural.png



Pre-WW2 U02 ceramic glaze.
Geo


Re: Testing for Cd in Zn Ore.

 

Next a scan of the empty chamber, same parameters but without a sample inside.
CD-Ka-La-xrf-test-307s-37kVp-25uA-NO-SAMPLE-IN-CHAMBER-small.jpg

Then the empty chamber "background with X-Ray ON" with the actual metal XRF together:
RED is the empty chamber with X-Ray ON, the black ghost outline is same parameters but with the pure metal "stamp" in place. Obviously the Cd L lines can be excited and measured in a pure sample. What remains next is to see how well they stand out in an actual ore. Right now I am of a mind that looking for Cd in Zn ore will best be serviced with an isotopic rather than electron exciter source.

Cd-Pure-Metal-Ghost-X-Ray-Source-Only-NO-SAMPLE-RED-BAK.png

Note the blue highlight is an ROI to indicate the area tat Cd L-lines occupy.


Geo










Testing for Cd in Zn Ore.

 

Thanks to member Steve D. in ALBQ NM for mentioning that Zn ores are also a common source of?
Cd.

Testing several species of ore here with the new emphasis on Cd has proved problematic. First the X-Ray hump from the tube is most intense right where Cd K line XRF is located, making it harder to pick out a weak peak there. Strong K peaks- no problem, as shown in one of the pictures to follow.

So I tried looking for L peaks from Cd in? ore.?

First steps are to use pure Cd metal to establish calibration points exactly for 3.2, 3.13, 3.31, 3.52, and 3.71kev.

Then another XRF test of the empty chamber, with no samples inside, for the same time and exposure data.

In all tests for this series, the 3 Al beam filters remain but the final collimator is removed to give a wide beam on the mineral tests to follow. The detector is an Amptek Si-PIN into the Amptek DP5-PC5 electronics stack, then USB to a notebook computer for control and storage.

First the pure metal "stamp"r, X-Ray source set to 37kVp-25uA

CD-Pure-Metal-Stamp-Ka-La-xrf-test-37kVp-25uA-307s-Notes1-small.jpg


In the above, Ka, Kb are in perfect alignment and something is barely detected around 3 keV This rendition (LINEAR) shows how well the K lines are excited in the lighter elements, compared to their L lines.

Geo


Re: Smithsonite with Uranium

 

开云体育

Cd commonly found in zinc ores, in this case probably present as the cadmium carbonate otavite. Zinc ores are the primary source of cadmium. Under UV, it has a very nice red fluorescence.


From: [email protected] <[email protected]> on behalf of Dude <dfemer@...>
Sent: Sunday, February 16, 2020 12:36 PM
To: [email protected] <[email protected]>
Subject: Re: [XRF] Smithsonite with Uranium
?

Geo,

Big difference. Zn predominant some Fe, Pb, Cd (!) ?and possibly Cu. Where’s the Uranium?

Dud

?

Smithsonite 90 sec si-pin dfe.bmp

?

?

?

?

?

?

From: [email protected] [mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Sunday, February 16, 2020 9:58 AM
To: [email protected]
Subject: Re: [XRF] Smithsonite with Uranium

?

On Sun, Feb 16, 2020 at 10:54 AM, Dude wrote:

One could hardly call that a peak. This is going to need the x-ray source to see what’s there.

Dud


This was cool. It was still running in the Gamma Spec chamber on the CdTe (no external exciter) so I stopped that count, moved it over to the XRF chamber and did a scan there with X-Ray tube and Si-PIN. That didn't take long, now the sample is back in the CdTe Gamma Spec chamber with the counting resumed. It's se for 24 hours, and will be finished shortly.?

If anyone is familiar with CZT detectors (Cadmium-Zinc-Telluride), Amptek tried them for a short time a maybe 15 years ago. Their performance is excellent, but there were problems with the pulse shape ( I think). They soon switched to a better all around crystal, this one is just Cadmium-Telluride evidently.
That's what is in the one chamber, we call it CdTe. It has good performance but not as good as Si-PIN, however, the CdTe carries that good performance up to much higher energies. It has lost no sensitivity at 100 keV and is very usable at 200 keV, and will go up to 400 keV max with my setup.


In the XRF chamber is a Si-PIN detector, it is only to specs up to 25 keV and falls off the charts at 60 keV.

Both technologies together give make a good combination I think.

Now the XRF scan, note the scan time!

Smithsonite_XRF_Graph.jpg

Also are attached 2 .mca, the 5 second one for fun and the serious 90 second one for hidden details.

Geo>K0FF

?


Re: UX2-234: Testing this unusual 1" beta check disc.

 

And UX1 and Ux2 are?

Say, you used ROI's for that display. Can you get PEAK SEARCH to do as well?

So far I haven't mastered Peak search and don't trust it.

Geo

----- Original Message -----
From: Dude <dfemer@...>
To: [email protected]
Sent: Sun, 16 Feb 2020 15:01:13 -0500 (EST)
Subject: Re: [XRF] UX2-234: Testing this unusual 1" beta check disc.

UX2 is a Uranium derived beta source used with a UX1 disk to determine
the solid angle correction factors for an end window ?beta counter.

Dud

From: [email protected]
[mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Sunday, February 16, 2020 10:40 AM
To: [email protected]
Subject: Re: [XRF] UX2-234: Testing this unusual 1" beta check
disc.

?

Anyone familiar with UX2?


Geo







Re: UX2-234: Testing this unusual 1" beta check disc.

 

开云体育

UX2 is a Uranium derived beta source used with a UX1 disk to determine the solid angle correction factors for an end window ?beta counter.

Dud

From: [email protected] [mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Sunday, February 16, 2020 10:40 AM
To: [email protected]
Subject: Re: [XRF] UX2-234: Testing this unusual 1" beta check disc.

?

Anyone familiar with UX2?

Geo


Re: Smithsonite with Uranium

 

开云体育

Geo,

Big difference. Zn predominant some Fe, Pb, Cd (!) ?and possibly Cu. Where’s the Uranium?

Dud

?

Smithsonite 90 sec si-pin dfe.bmp

?

?

?

?

?

?

From: [email protected] [mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Sunday, February 16, 2020 9:58 AM
To: [email protected]
Subject: Re: [XRF] Smithsonite with Uranium

?

On Sun, Feb 16, 2020 at 10:54 AM, Dude wrote:

One could hardly call that a peak. This is going to need the x-ray source to see what’s there.

Dud


This was cool. It was still running in the Gamma Spec chamber on the CdTe (no external exciter) so I stopped that count, moved it over to the XRF chamber and did a scan there with X-Ray tube and Si-PIN. That didn't take long, now the sample is back in the CdTe Gamma Spec chamber with the counting resumed. It's se for 24 hours, and will be finished shortly.?

If anyone is familiar with CZT detectors (Cadmium-Zinc-Telluride), Amptek tried them for a short time a maybe 15 years ago. Their performance is excellent, but there were problems with the pulse shape ( I think). They soon switched to a better all around crystal, this one is just Cadmium-Telluride evidently.
That's what is in the one chamber, we call it CdTe. It has good performance but not as good as Si-PIN, however, the CdTe carries that good performance up to much higher energies. It has lost no sensitivity at 100 keV and is very usable at 200 keV, and will go up to 400 keV max with my setup.


In the XRF chamber is a Si-PIN detector, it is only to specs up to 25 keV and falls off the charts at 60 keV.

Both technologies together give make a good combination I think.

Now the XRF scan, note the scan time!

Smithsonite_XRF_Graph.jpg

Also are attached 2 .mca, the 5 second one for fun and the serious 90 second one for hidden details.

Geo>K0FF

?


Re: UX2-234: Testing this unusual 1" beta check disc.

 

Anyone familiar with UX2?

Geo


Re: Smithsonite with Uranium

 

On Sun, Feb 16, 2020 at 10:54 AM, Dude wrote:

One could hardly call that a peak. This is going to need the x-ray source to see what’s there.

Dud


This was cool. It was still running in the Gamma Spec chamber on the CdTe (no external exciter) so I stopped that count, moved it over to the XRF chamber and did a scan there with X-Ray tube and Si-PIN. That didn't take long, now the sample is back in the CdTe Gamma Spec chamber with the counting resumed. It's se for 24 hours, and will be finished shortly.?

If anyone is familiar with CZT detectors (Cadmium-Zinc-Telluride), Amptek tried them for a short time a maybe 15 years ago. Their performance is excellent, but there were problems with the pulse shape ( I think). They soon switched to a better all around crystal, this one is just Cadmium-Telluride evidently.
That's what is in the one chamber, we call it CdTe. It has good performance but not as good as Si-PIN, however, the CdTe carries that good performance up to much higher energies. It has lost no sensitivity at 100 keV and is very usable at 200 keV, and will go up to 400 keV max with my setup.


In the XRF chamber is a Si-PIN detector, it is only to specs up to 25 keV and falls off the charts at 60 keV.

Both technologies together give make a good combination I think.

Now the XRF scan, note the scan time!

Smithsonite_XRF_Graph.jpg


Also are attached 2 .mca, the 5 second one for fun and the serious 90 second one for hidden details.

Geo>K0FF

?