Re: Tek 577 unable to display more than 5 steps
Thank you for your help, Ozan,
I have posted a picture of the signals at Q310 base & collector and Q310 base & emitter on the photo album?/g/TekScopes2/album?id=279289. The base signal really follows the emitter, but the voltage difference seems to be 7.5V rather than 0.75V. This time I am using two tek probes. Am I reading the voltage difference with the cursors wrong? If not, what may be causing this? |
Re: SC 503 Enhance/Integrate removed on later units
Hi Stefan,
Look at the photo directory, I've added a screenshot of the older schematic along with photos. For my restauration, I'll eventually have to buy the SN B040000 manual from Artek Media, but right now I only have access to older one from tekwiki. The newer version has green PCB and missing components. So my guess the PCB is basically the same, minus the components of course. Note that A20 board (focus & intensity) has already 4 revisions? (part 670-5369-0[0-3]) for the older version. It would be interesting to look at the part number of the newer version. Also, may be they use a different tube (V500), on the older one, this is part 154-0812-00. Best regards, |
Re: SC 503 Enhance/Integrate removed on later units
Ke-Fong Lin,
?
please take a closer look, if some components are not installed in the newer one. (Should be according to newer manual.) The F&I BD is on the left side. I've added a screenshot schematic 10 for B040000-up:
/g/TekScopes2/album?id=280676
The newer manual only states 50cm/s stored writing speed. Old version was 250!
The newer manual can be obtained from Artek Media.
?
Stefan |
Re: Tek 577 unable to display more than 5 steps
On Sun, Oct 23, 2022 at 01:34 PM, Rogerio O wrote:
Hi folks,¡ª¡ª¡ª- Hi Roger, ? S429 (PULSED mode) switch should be pulling cathode of CR234 to ground in normal mode, which should pull base of Q234 below turn on voltage (Vbe below 0.5V). With Q234 off R234 should pull base of Q320 high. Base of Q310 should follow the ramp with a 0.7V offset.? ? If you pull out Q234 do you see all the steps? If it works possible failure modes in the order of probability: dirty PULSED switch, open CR234, open or drifted R237, bad Q234, drifted R234.? ? Since you already replaced Q310 we can assume it is good.? Ozan |
Re: SC 503 Enhance/Integrate removed on later units
Hi Stefan,
I've actually both a SN B040000 and up newer version, and an older one. They're both non-working and restauration projects. However, I've taken a quick look already, and if I remember correctly, the PCB are actually same. But with the enhance/integrate switch/potentiometer combo not installed (holes are there!).?Otherwise components seem same. I can have closer look if you want. BTW, do you have the newer version's manual? Tekwiki only has the older one. Best regards, |
Re: Tek 577 unable to display more than 5 steps
¿ªÔÆÌåÓýIF Q310 is used as a current amplifier (low impedance input) then the base voltage can be 0.6 volts even if the operating conditions are bad.? Is the base connected to multiple sources through resistors?? If so, it is likely a current amplifier.? Please check the voltages and signals at the other end of such resistors.? Check the voltage drops across the collector and emitter resistors to see if they are correct, likewise, too low a collector voltage supply can cause saturation.? Try these tests with no steps generated (if possible) to see the DC operating points. Harvey
On 10/23/2022 4:34 PM, Rogerio O via
groups.io wrote:
Hi folks, |
Re: Tek 577 unable to display more than 5 steps
Hi folks,
Following Ozan's suggestion I have measured pin7 in on CH1 and pins 5 and 6 on CH2 and everything went as expected. 0V at pins 5 and 6 while having the staircase signal at pin 7. Then I decided to replace Q310 with an originial tek 151-0192-00 that is used in the later models. Since the problem was still there I measured the signals at the collector of Q310 (TP305) and its emitter (left side of R310) and, to my surprise the signal was "saturating" there. I posted a picture of the signals st the photo album?/g/TekScopes2/album?id=279289.?? PLEASE NOTE THAT THE PROBE ON CH2 IS NON TEK SO THE REAL V/Div READING IS 2V/DIV!!! So I checked the DC bias at Q310??base (using the settings of table 7-2)? and the mesured +0.6V as shown in schematics. I made a "crude verification of signal loading" by measuring the resistance to ground of U350B pin 6? (13.4K) and pin7 (5K), but I don't know if these value may cause (or explain) the problem. Any other suggestion? Thanks ? |
AFG5101, arb func gen, looking for manuals....?
Hi all,
I am currently looking for copies or scans of the Instrument Interfacing Guide (part no. 070-6759-00) and the Reference Guide (part no. 070-6761-00) for the AFG5101 arbitrary function generator. These seem to be a bit rare, so the usual suppliers (Artek etc) don't seem to have them... Anyone got a copy of these they can send/sell to me by chance? Thanks! Jared |
SC 503 Enhance/Integrate removed on later units
The Enhance/Integrate knob + corresponding electronic components (Storage schematic 10) have been removed on later units, SN B040000 and higher. The remaining circuit + CRT was mainly unchanged.
Why has this been done? Did it stress the CRT too much?
?
Stefan |
Re: First post here...
Bonjour ch¨¨re Jennifer,
We have met before..am sure you knew our friends in audio....and Tek... Sid S. Dick S. (RIP), Rich C, Bruce H Tomlinson H Sol Marantz Bravo for the Tektronix collection! Fine selection. See the other Tekscope at groups.io ( omit the 2) as they do not overlap, also Tekwiki and EEVBLOG. Bon Soir¨¦e Jon |
Re: First post here...
¿ªÔÆÌåÓýJennifer- Welcome aboard......and here I thought I had too many...space is the limiting factor.....I am not a collector either but grew up using mostly Tek scopes. Great bunch of folks here. ¸é±ð²Ô¨¦±ð On 10/22/22 1:10 AM, Jennifer Crock via
groups.io wrote:
Greetings. I work in the audio industry and had my teen years in the back yard of Tektronix , so naturally gravitated to using Tek gear. I am not a collector, just a user,... I use my Tek stuff in a real working environment. Here is a partial list of my lab: Jennifer <7000> 7104 opt 2 & |
Re: First post here...
John Griessen
On 10/22/22 02:10, Jennifer Crock via groups.io wrote:
Here is a partial list of my lab:I like the 7844 and 7104 also. Welcome to a place to ask questions about repair, use of probes, measurement setups, whatever. I have spare scopes untested also. I am not a collector either. I only use tektronix scopes for very serious applications -- never because they can be sort of beautiful. I also try to spend more time measuring with them than repairing or polishing. Not a collector. :-) -- John Griessen -- building lab gear for biologists Albuquerque NM blog.kitmatic.com 7104 opt 2 & 3 1 Gig |
Re: First post here...
hmmm, not a collector........... On Sat, Oct 22, 2022 at 4:42 AM Jennifer Crock via <collies13=[email protected]> wrote: Greetings. |
First post here...
Greetings.
I work in the audio industry and had my teen years in the back yard of Tektronix , so naturally gravitated to using Tek gear. I am not a collector, just a user,... I use my Tek stuff in a real working environment. Here is a partial list of my lab: Jennifer <7000> 7104 opt 2 & 3 1 Gig 7A29 1 GHz 50ohm single 7A19 600MHz 50ohm single 7B10 1 GHz 200mS to 2 nS x10=200pS 7B15 1 GHz Delaying " " " R7844Dual Beam 400MHz scope 7A24 2 ch 400MHz 7A16A single 150MHz 7B92A 500MHz time base 7B92A 500MHz time base 7834 400 MHz Storage scope 7A16 single 150MHz 7A18 2 ch 75MHz 1meg 7B92A 500MHz time base 7B92A 500MHz time base Untested, no plug-ins installed 7704A / P7001 digitizer scope Extra 7000 series items 7A13 Diff Comparator 100MHz 7A13 Diff Comparator 100MHz 7A13 Diff Comparator 100MHz 7A17 150MHz high gain 7A18 2 ch 75MHz 1meg 7B53N 100MHz time base 7B92 500MHz time base 7D10 event count to trigger 7D11 delay time to trigger 7D14 freq counter Other series 2247A 100MHz Portable 576 Curve Tracer 576 Curve Tracer 575 Curve Tracer 535 10MHz Scope Also use a pretty good collection of TM500 and 5000 items including Bruce Hoffer's audio masterpieces, the AA501A and SG505 with IMD and balanced out options. <TM 500 & 5000> TM5000 w/rack ears TM5000 w/rack ears TM506 rack TM506 rack TM506 rack TM506 rack TM515 suitcase TM501 TM501 TM503 TM504 TM504 AA501 Distortion test AA501A Distortion test AM503 current probe set with A6302 Current probe DM501 4.5 digit DMM DM501A 4.5 digit DMM DM501A 4.5 digit DMM DM501A 4.5 digit DMM DM501A 4.5 digit DMM DM501A 4.5 digit DMM SG502 5Hz 500KHz SG502 5Hz 500KHz SG502 5Hz 500KHz SG503 leveled sine 250MHz SG505 opt 2, balanced out, IMD FG504 sweep gen 40MHzl FG507 sweep gen .002 to 2MHz PG502 pulse 10Hz to 250MHz PG505 pulse 1Hz to 100KHz TG501 time base gen DC503A counter DC505A counter PS503A dual power supply PS501-1 PS lab-built dual supply SC504 40 meg scope PS5010 power supply PS5010 power supply DM5010 multi meter DM5010 multi meter FG5010 20meg function FG5010 20meg function 2 ) probe tool case plug in 5 ) slot covers |
Re: AFG5101 Option 02, can someone check a part number on theirs?
Guess I should have remembered to look there first...
As a last resort, you may need to look at the schematics from some other signal generators of that era and look for the same looking pin-out device. See if there is a part number for that. Transistors might just be general purpose, and crystal footprint generic so should not stop the design of a board, like the ICs would. Someone will eventually pop up with the values. -- T. Gerbic Central California |