Re: Tektronix 465,NO Trigger

 

Hi John,
No problem about the time... It's a relief for my mind to solve those puzzles.
Some comments further down, just after your answers...
Regards,
Fabio

On Fri, Aug 17, 2018 at 07:16 AM, John Stoole wrote:
Hello Fabio , I decided to contact you direct in case I want to send you any photos of my problems here,
Firstly I want to thank you for the very detailed help you sent me, it must have taken up a lot your your valuable time and it is much appreciated.
I have now completed all the suggestions that you made with no results but here are some of the things that I found
1. A trigger level control has no effect with any of the switch positions
3. Pressing the TRIG VIEW I have a blank screen
Hmmm... Both things above together, may mean that the triggering signal, through the triggering input and pre-amplifier stages may be so much offset in D.C. that the triggering signal coming out of the pre-amplifier into the triggering discriminator (Tunnel Diodes or Tek Custom IC yet to be clarified) is never crossing the 0V (which is a must, for the trig. discriminator to operate).
**Please try this one more thing**
While pressing the Trig View, press also the Beam Finder (you should see the trace either stuck near the top or the bottom of the screen).
While keeping both pressed, move the A Trigger Level, to see if the trace shows any tendency to move towards the center of the screen...
This will give you some hints if the trigger preamplifier is really stuck at some DC level, or if it still reacts to the A trig level pot.

2. If I switch the A COUPLING to DC, then as I step the SOURCE switch from NORM to EXT10 the trigger lamp flashes proving the lamp is OK
This is good! It gives us some indication that the triggering discriminator is working... It's probably detecting the triggering signal crossing 0V, when you're changing selections, which is probably causing some change in the DC levels at the input stages...

4. the TRIG HOLD OFF control works well and helps to lock the waveform when moved
Expected... The Sweep control logic seems to be working, otherwise there wouldn't be a trace whatsoever.

5 Tried 5 v AC into the external trigger but that would not sync the waveform
This indicates that whatever the problem is, it's probably at or after the trigger pre-amplifier input selector (e.g. it sort of rules-out that your problem is at the pick-off circuitry).
In time, the pick-off circuitry are the pieces of circuitry of the Vertical pre-amplifiers that "taps" the vertical signal from Channel 1, Channel 2 or from after the channel Alternating circuit (the NORM pick-off) and routes it to the trigger preamp.
If the problem would be at the pick-off circuitry (only), you would still be able to trigger from an EXT input.

the rest of the scope seems to work very well
I will continue to hunt around the the voltages to see if I can find any serious deviations for the service manual voltages
To further help you from this point on, we will need to know:
A. What kind of Trigger pre-amp board your 465 has. I`m not even sure if this "doubt" applies to the 465, but I used to have a 464 and I know that (for the 464 at least) there were two versions of Trigger boards, one that uses Tunnel Diodes, and another one that uses a Tek custom IC (as the triggering signal discriminator).

Many thanks again for the help
regards
John


Re: Tsk 2467b test 05 error

 

On Fri, Aug 17, 2018 at 04:38 <Sscandizzo@...> wrote:

Hi Everyone,

I'm definitely a newb to electronics repair, so I wanted to get some
advice about my Tektronix 2467b. On startup, the scope errors out with one
of two codes: "Test 05 Fail 04" or "Test 05 Fail 42". However, if I run
the diagnostics (single time or loop), test 5 passes.

Generally, the scope seems to be behaving itself (nothing obvious).

Any suggestions on where to start or why it is behaving like this?

Hey Stefan,

This is typically because of leaking surface mount electrolytic capacitors
on the A5 board. If you consult the archives, you¡¯ll find plenty of
information about this problem. Those capacitors will need replacing and
the PCB cleaned to prevent further damage.
The error occurs because the electrolyte wrecks the resistors (or PCB
tracks) that provide the reference for the DAC, so you¡¯ll have to see about
replacing those and/or repairing the PCB. The A5 board generates two
reference voltages, 1.36V and -1.25V. If you measure those, I bet you¡¯ll
find them off a bit.
Note that at least one of those resistors is a low temperature coefficient
type, which helps with holding calibration and working in spec across the
operating temp range.

The other possibility is that there is something going on with the line
triggering path, but even then you¡¯d want to replace those capacitors
preemptively.

Good luck,
Siggi


Re: Tektronix 465,NO Trigger

John Stoole
 

Thanks for the very comprehensive reply Fabio I will start to carry out the suggestions you have made, and will let you know how I get on
73
John

On 16 August 2018 at 22:02 Fabio Trevisan <fabio.tr3visan@...> wrote:


Hello John,
As a general rule of thumb, a thorough check of the Low Voltage power supplies, both for voltage on spec and ripple are both important and yet, an easy check.
Not just because problems at the power supplies may cause, at times, unpredictable problems that are hard to make sense of (even odd ones such as everything working BUT syncing), but also because finding an off-spec power supply that has not yet caused a serious damage, can save you from watching the inevitable magic smoke, and the need for a much more complicated repair.
Having made sure the power supplies are healthy, the next step (in your case, where the scope seems to be working to a great extent) is to get acquainted with the operation of your scope, to be sure that the position of the controls are all correct for the operation that you're attempting and for the signals you're applied.

More specific to the symptom you mentioned, here are the basics...
1. Exercise all selector levers, push-buttons and pots a dozen or more times... to make sure that you cleared any false contact that may be present, due to oxidation of contacts, or dried lubricant crud.
2. Make sure Horizontal Display mode ganged push buttons is set to A sweep (button "A Lock Knobs" pressed),
3. That the A and B Time/Div dials are locked (that both the clear dial skirt and the gray knob are aligned and moving together) and a suitable sweep speed is selected.
4. Trigger mode ganged push buttons is set to AUTO (button AUTO pressed).
5. That the A Trigger level pot is set to middle range.
6. Vertical signal level has some healthy 4 or 5 graticules in height (and that you use only one channel (for now)
7. A trigger coupling lever is set to AC
8. A trigger source lever is set to NORM.

From this starting point, if it doesn't sync right away, then there must be some issue at play... and it will become a matter of narrowing it down, by fiddling with the controls, one at a time, in a combinatory fashion... For instance:
A. From the starting point above, try to move (slowly) the A trigger level pot, from CCW to CW. If no avail, put if back to center.
B. Try to change the Source lever, from NORM, to the specific channel where you have your vertical signal fed to (CH1 or CH2), and repeat sweeping the A trig level from full CCW to CW.
C. If still no avail, change the Triggering coupling from AC, to HF REJ, to LF REJ and DC... for each possible coupling, try the steps A and B above.
D. If neither work, use a BNC "tee" and a coax cable, and try to feed the same vertical signal to the A EXT input, in combination with changing the the Source lever to "EXT" or "EXT¡Â10". At this step, try playing with all combinations of the Coupling lever while sweeping the A trigger level pot from CCW to CW.
E. If neither work, then you can still try to feed a vertical signal from the AC mains (or the output of an AC transformer), select an appropriate sweep speed and change the Source lever to "LINE". While at this step, exercise again the Trigger coupling lever and the A trig level potentiometer...

As an extra resource of information, while at each of the lettered steps above, you can try to press the trig view switch.
Under normal conditions, on a working oscilloscope... while pressing the TRIG VIEW push button, you should see on the screen (instead of your Vertical signal), a "fac simile" of the triggering signal that is coming out from the A trigger amplifier, past all the coupling, source and level controls....It should **normally** display much larger on the screen than your original vertical signal is showing.
While you press the TRIG VIEW button, rotating the A TRIG Level potentiometer should make the signal cross the screen from top to bottom (while turning the A trig level from CCW to CW).
If you see the trigger signal while pressing the TRIG VIEW button, and if it reacts to the A trig level control, and the scope still doesn't sync, then you have a problem on the triggering circuitry past the triggering pre-amplifiers.
If you don't see the trigger signal while pressing the TRIG VIEW button, or it stays stuck at either the top or the bottom of the screen, then you have a problem on the trigger input circuitry or the trigger pre-amplifier.

Hope this helps you move further...

Rgrds,

Fabio

On Thu, Aug 16, 2018 at 05:38 PM, <john@...> wrote:


Greetings All, I recently purchased a 465 on Ebay UK, having now cleaned it up
it's looking good, It seems to be working with a good sine wave on both
channels from the calibrator, but the only way I can trigger the waveform is
by very careful adjustment on the TIME/DIV control and the little red Knob in
the center, I don't have a Trigger light ( the lamp in OK) so I presume that I
have a Trigger fault. I have checked all the voltages and all are in spec.
Although I am fine when repairing Ham Transceivers I know nothing about
Scopes.
Can any kind expert give me some suggestions as to what to test first ??
Thanking you in advance
John
mi0dfg


Re: Tektronix 465,NO Trigger

 

Thanks Harvey, all suggestions are very helpful as I said before I am useless when it comes to scopes and it a big learning curve!!!!
From all the tests I have made there seems to be a complete lack of any trigger action, and No trigger lamp indication, will carry out the tests you have indicated and once again thanks for taking the time to offer help
73
John

On 17 August 2018 at 00:18 Harvey White <madyn@...> wrote:


On Thu, 16 Aug 2018 10:18:01 -0700, you wrote:

Greetings All, I recently purchased a 465 on Ebay UK, having now cleaned it up it's looking good, It seems to be working with a good sine wave on both channels from the calibrator, but the only way I can trigger the waveform is by very careful adjustment on the TIME/DIV control and the little red Knob in the center, I don't have a Trigger light ( the lamp in OK) so I presume that I have a Trigger fault. I have checked all the voltages and all are in spec.
Although I am fine when repairing Ham Transceivers I know nothing about Scopes.
Triggers are derived from the appropriate channel, or externally.
They're processed, and that signal is used to trigger the sweep.

AUTO triggers the sweep at a fixed rate in the absence of a trigger.
If it works on auto, then the sweep should be able to be triggered.

That you can trigger, but only at a specific spot, suggests that the
trigger amplitude is *really* small somewhere.

This is a case where you'd love to have a scope to test the scope.

I'd get the manual, then look at where the signal comes from the
vertical amplifier (there's a switch in there). Start tracing the
signal from the vertical amplifier through the trigger circuits,
observing voltages and voltage limits on controls.

If everything is OK, and you get the right signals, then check the
input to the sweep generator, that might be bad.

Harvey


Can any kind expert give me some suggestions as to what to test first ??
Thanking you in advance
John
mi0dfg





Tsk 2467b test 05 error

 

Hi Everyone,

I'm definitely a newb to electronics repair, so I wanted to get some advice about my Tektronix 2467b. On startup, the scope errors out with one of two codes: "Test 05 Fail 04" or "Test 05 Fail 42". However, if I run the diagnostics (single time or loop), test 5 passes.

Generally, the scope seems to be behaving itself (nothing obvious).

Any suggestions on where to start or why it is behaving like this?

Thanks!
Stefan


466 power transistor

 

I have to replace the large power transistor on the A6 inter board the TIP35A one it was damaged from prior ill handling plus it appears to have been overheated at sometime as the case is deformed. I found a replacement on Mouser and have ordered it not expensive so I'm asking if ya'all think it will work as well as the original? It is a Central made & supplied version and it is now full plastic case no metal tab it is like the 475 ones. It is rated 25A, 60V 125W don't know what original is as no specs in manual.
Also I have discovered that the A6 board has been updated to the later B2000000 and on type parts added.
My scope sn#168*** and my board is -09 not listed in either manual?
When i get it together and calibrating I will be here for advise on proper adjustments.
On the A10 storage board cap C1967 is a blue Sprague 1uf 150v aluminum lytic not a wet tant, interesting the other two are still wet tants.

Jim


Re: 453 trouble

 

Hello Guys,


Many thanks for technicals comments, thats great !

I must keep your advice in a corner of my brain when I work

on any stuff with digital signal.

Regards

Alain

Le 16.08.2018 ¨¤ 15:35, Harvey White a ¨¦crit?:
On Thu, 16 Aug 2018 06:05:57 +0200, you wrote:

Hello Folks,


Thanks for all thoses interestings comments.

As Ive reconnected the ground, and turned power ON : no differential
main shut down occur, even when my finger

played with chassis parts. I also mesured voltage between ground and
chassis : close to zero -- few mV-..

The idea to use the same wall socket to avoid ground loop is fine.

Question : sometime when troubleshooting radios, I use a ground lead
between scope and radio to avoid using probe's aligator clip.

What do you think about ? can I introduce false mesurements ?
Yes and no. The ground lead for the probe, when properly attached,
minimizes ringing on the signal. The long lead from chassis to
chassis will not. For low frequency signals (especially sine waves),
you're not likely to see much of a difference. Go play digital, and
you will. The length of the probe's ground lead and where it is
placed also have an effect, especially if you've got a different
ground potential (by millivolts) across a PC board or a chassis.

Harvey



Regards

Alain




Le 15.08.2018 ¨¤ 20:40, Richard Knoppow a ¨¦crit?:
??? A hot chassis can also be caused by RF filter capacitors connected
from each side of the line to the chassis. I would be much more
suspicious of these if they exist in the 453 than of the electrostatic
shield in the power transformer. If there are such caps in the 453
lift them to see if the voltage on the chassis disappears. You might
want to replace them. There are special capacitors for this service
which fail open so they don't connect the line directly to the chassis.
??? The voltage on the chassis will depend on the amount of
capacitance and the impedance of the measuring device.
??? You can always generate a long argument by discussing the
differences between lightening protection grounds, power line safety
grounds and grounds to eliminate common impedances in AF and RF circuits.

On 8/15/2018 1:52 AM, Craig Sawyers wrote:
That would be typical if the chassis is ungrounded. Capacitance
between the primary and shield (which
is connected to the chassis) in the mains transformer will cause
this. First thing to do is connect
the ground in the mains plug - the previous user did a dangerous
thing by disconnecting the ground.

There may be other problems to solve, of course - but make the scope
safe first.

Craig




Re: 422 AC power socket

 

On Tue, Aug 14, 2018 at 04:53 PM, <johnasolecki@...> wrote:


10 seconds with my pocket knife and I now have a 50' power cord!
Fifty feet ? Nothing, I got batteries !

LOL

But if I ever need a cord I have no idea where to get one and I don't even have the DC cord. I reluctantly recommend a modification - to myself. Hell, it's only 2 wires.


2215 Problems

 

Hi all, Just found out about this group and I hope you all can help me out a little. I bought a 2215 off ebay (and subsequently got my $40 back no questions asked, so not only did he know it was broken despite listing as used, free scope!). Anyways, I started to troubleshoot using the service manual, since the scope just hummed and no trace was provided on the screen without any amount of fiddling, and I tested the voltages on each of the test points and none were anywhere near the acceptable range (ex. the 30v rail was at 5v...) So I am assuming this is probably caused by bad caps in the psu? Any further thoughts or troubleshooting steps are appreciated.


Re: 7704A Z-Axis Problem

 

Whoa, why did that cap blow in the first place?

I used my 7A22 and I remember that I was careful to keep the overdrive lamp off. I used the filters to remove LF and HF as much as I could so I like to think that my measurements were correct but I can't be certain.

I'll probably try to get a replacement HV Board just in case. It's making all kind of weird noises so I wouldn't be surprised if quit any time soon.

Anyone with a spare HV Board?

Thanks for the help!


Re: Tektronix 465,NO Trigger

 

"It seems to be working with a good sine wave on both channels from the calibrator"

???

Is this the calibrator on the scope? If so, it should be a nice square wave. Are you using a scope probe?

Regards

----- Original Message -----
From: <john@...>
To: <[email protected]>
Sent: Thursday, August 16, 2018 1:18 PM
Subject: [TekScopes] Tektronix 465,NO Trigger


Greetings All, I recently purchased a 465 on Ebay UK, having now cleaned it up it's looking good, It seems to be working with a good sine wave on both channels from the calibrator, but the only way I can trigger the waveform is by very careful adjustment on the TIME/DIV control and the little red Knob in the center, I don't have a Trigger light ( the lamp in OK) so I presume that I have a Trigger fault. I have checked all the voltages and all are in spec.
Although I am fine when repairing Ham Transceivers I know nothing about Scopes.
Can any kind expert give me some suggestions as to what to test first ??
Thanking you in advance
John
mi0dfg


Re: Tektronix 465,NO Trigger

 

On Thu, 16 Aug 2018 10:18:01 -0700, you wrote:

Greetings All, I recently purchased a 465 on Ebay UK, having now cleaned it up it's looking good, It seems to be working with a good sine wave on both channels from the calibrator, but the only way I can trigger the waveform is by very careful adjustment on the TIME/DIV control and the little red Knob in the center, I don't have a Trigger light ( the lamp in OK) so I presume that I have a Trigger fault. I have checked all the voltages and all are in spec.
Although I am fine when repairing Ham Transceivers I know nothing about Scopes.
Triggers are derived from the appropriate channel, or externally.
They're processed, and that signal is used to trigger the sweep.

AUTO triggers the sweep at a fixed rate in the absence of a trigger.
If it works on auto, then the sweep should be able to be triggered.

That you can trigger, but only at a specific spot, suggests that the
trigger amplitude is *really* small somewhere.

This is a case where you'd love to have a scope to test the scope.

I'd get the manual, then look at where the signal comes from the
vertical amplifier (there's a switch in there). Start tracing the
signal from the vertical amplifier through the trigger circuits,
observing voltages and voltage limits on controls.

If everything is OK, and you get the right signals, then check the
input to the sweep generator, that might be bad.

Harvey


Can any kind expert give me some suggestions as to what to test first ??
Thanking you in advance
John
mi0dfg



Re: 2215A LVPS repair

tom jobe
 

Hi Leo,
For some reason the photos you think you are adding to your photo album today, are not showing up in the photo album!
tom jobe...

On 8/16/2018 2:32 PM, satbeginner wrote:
Hi Tom,

Thanks for the documents, I added them to my 2215A paperwork and checked my scope for these mod's.

I am in Spain, but based on the SN (204xxx) my 2215A was manufactured in the UK, so it's probably always been in a 220V environment.
I got the impression these mod's were based on the early US build SN's, depending on scope model. (01xxxx - ), and not for the (later?) manufactured UK (2xxxxx) and Dutch (7xxxxx) SN's?
I checked my scope for these mod's, and some were already standard in the schematic of the 2215A, but funny enough there seem to be many iterations around of the PS design, some mod's are to be done only to 2235's, and some only to 2215 but not to 2215A, etc.

What I did do was:

Replaced the Schaffner line filter to be on the safe side;
Replaced the two (already micro cracked) 2n2 RIFA capacitors in the extra line EMI filter build to the back of the scope;
Kept the capacitor C900, it looks as if that one was replaced before;
Kept the power supply mod's as they are now.

Pictures of the removed components are here: /g/TekScopes/album?id=64919&p=pcreated,,,20,1,0,0

Thanks again for the input,

Leo


Re: Tektronix 465,NO Trigger

 

Hello John,
As a general rule of thumb, a thorough check of the Low Voltage power supplies, both for voltage on spec and ripple are both important and yet, an easy check.
Not just because problems at the power supplies may cause, at times, unpredictable problems that are hard to make sense of (even odd ones such as everything working BUT syncing), but also because finding an off-spec power supply that has not yet caused a serious damage, can save you from watching the inevitable magic smoke, and the need for a much more complicated repair.
Having made sure the power supplies are healthy, the next step (in your case, where the scope seems to be working to a great extent) is to get acquainted with the operation of your scope, to be sure that the position of the controls are all correct for the operation that you're attempting and for the signals you're applied.

More specific to the symptom you mentioned, here are the basics...
1. Exercise all selector levers, push-buttons and pots a dozen or more times... to make sure that you cleared any false contact that may be present, due to oxidation of contacts, or dried lubricant crud.
2. Make sure Horizontal Display mode ganged push buttons is set to A sweep (button "A Lock Knobs" pressed),
3. That the A and B Time/Div dials are locked (that both the clear dial skirt and the gray knob are aligned and moving together) and a suitable sweep speed is selected.
4. Trigger mode ganged push buttons is set to AUTO (button AUTO pressed).
5. That the A Trigger level pot is set to middle range.
6. Vertical signal level has some healthy 4 or 5 graticules in height (and that you use only one channel (for now)
7. A trigger coupling lever is set to AC
8. A trigger source lever is set to NORM.

From this starting point, if it doesn't sync right away, then there must be some issue at play... and it will become a matter of narrowing it down, by fiddling with the controls, one at a time, in a combinatory fashion... For instance:
A. From the starting point above, try to move (slowly) the A trigger level pot, from CCW to CW. If no avail, put if back to center.
B. Try to change the Source lever, from NORM, to the specific channel where you have your vertical signal fed to (CH1 or CH2), and repeat sweeping the A trig level from full CCW to CW.
C. If still no avail, change the Triggering coupling from AC, to HF REJ, to LF REJ and DC... for each possible coupling, try the steps A and B above.
D. If neither work, use a BNC "tee" and a coax cable, and try to feed the same vertical signal to the A EXT input, in combination with changing the the Source lever to "EXT" or "EXT¡Â10". At this step, try playing with all combinations of the Coupling lever while sweeping the A trigger level pot from CCW to CW.
E. If neither work, then you can still try to feed a vertical signal from the AC mains (or the output of an AC transformer), select an appropriate sweep speed and change the Source lever to "LINE". While at this step, exercise again the Trigger coupling lever and the A trig level potentiometer...

As an extra resource of information, while at each of the lettered steps above, you can try to press the trig view switch.
Under normal conditions, on a working oscilloscope... while pressing the TRIG VIEW push button, you should see on the screen (instead of your Vertical signal), a "fac simile" of the triggering signal that is coming out from the A trigger amplifier, past all the coupling, source and level controls....It should **normally** display much larger on the screen than your original vertical signal is showing.
While you press the TRIG VIEW button, rotating the A TRIG Level potentiometer should make the signal cross the screen from top to bottom (while turning the A trig level from CCW to CW).
If you see the trigger signal while pressing the TRIG VIEW button, and if it reacts to the A trig level control, and the scope still doesn't sync, then you have a problem on the triggering circuitry past the triggering pre-amplifiers.
If you don't see the trigger signal while pressing the TRIG VIEW button, or it stays stuck at either the top or the bottom of the screen, then you have a problem on the trigger input circuitry or the trigger pre-amplifier.

Hope this helps you move further...

Rgrds,

Fabio

On Thu, Aug 16, 2018 at 05:38 PM, <john@...> wrote:


Greetings All, I recently purchased a 465 on Ebay UK, having now cleaned it up
it's looking good, It seems to be working with a good sine wave on both
channels from the calibrator, but the only way I can trigger the waveform is
by very careful adjustment on the TIME/DIV control and the little red Knob in
the center, I don't have a Trigger light ( the lamp in OK) so I presume that I
have a Trigger fault. I have checked all the voltages and all are in spec.
Although I am fine when repairing Ham Transceivers I know nothing about
Scopes.
Can any kind expert give me some suggestions as to what to test first ??
Thanking you in advance
John
mi0dfg


Re: 2215A LVPS repair

 

Hi Tom,

Thanks for the documents, I added them to my 2215A paperwork and checked my scope for these mod's.

I am in Spain, but based on the SN (204xxx) my 2215A was manufactured in the UK, so it's probably always been in a 220V environment.
I got the impression these mod's were based on the early US build SN's, depending on scope model. (01xxxx - ), and not for the (later?) manufactured UK (2xxxxx) and Dutch (7xxxxx) SN's?
I checked my scope for these mod's, and some were already standard in the schematic of the 2215A, but funny enough there seem to be many iterations around of the PS design, some mod's are to be done only to 2235's, and some only to 2215 but not to 2215A, etc.

What I did do was:

Replaced the Schaffner line filter to be on the safe side;
Replaced the two (already micro cracked) 2n2 RIFA capacitors in the extra line EMI filter build to the back of the scope;
Kept the capacitor C900, it looks as if that one was replaced before;
Kept the power supply mod's as they are now.

Pictures of the removed components are here: /g/TekScopes/album?id=64919&p=pcreated,,,20,1,0,0

Thanks again for the input,

Leo


Re: 2215A LVPS repair [Solved]

 

On Thu, Aug 16, 2018 at 10:55 PM, Bert Haskins wrote:




On 8/16/2018 9:43 AM, tom jobe wrote:
Hi Leo,
Thanks for the progress report, It's good to see that your 2215A
repairs so far have gone well!
In studying your photos, it looks like you have one of those dreaded
"Schaffner" brand AC inlet filters (where the AC cord plugs into the
back of the scope).
If that is what you have, they have a poor reliability record when
used on 240 AC line voltage, if you also have that too.
I'm in a 120 volt AC part of the world and I started changing them
after having just one scary "Schaffner" failure.
The Tekscopes message archive should have some "Schaffner" discussions
from the past, and a Google search on the term 'Schaffner filter
failure' will get you more information too.
You might also consider changing the few X and Y caps in the 2215A,
especially C904 which is under the AC inlet filter and behind the main
power switch.
C904 is much easier to change while you have the AC inlet filter removed.
The original X and Y capacitors have difficulty with 240 volt mains,
so you might want to get the highest rated replacements you can find.
tom jobe...

On 8/16/2018 4:19 AM, satbeginner wrote:
Hi all,

just a final message: it's all done!

Today I received the 12V 60mm brushless fan that was a perfect fit.
I used the half wave rectifier circuit I was able to place the
components for on the board, the fan is running nicely without real
noise on 9V.
I mean, you can hear it, but only just.
There are plenty perforations all around the casing to let cool air
in, so I choose to have it blowing out, so it is sucking the warm air
from the inside of the scope and remove it from the scope.
(Like in my 2465's)

One extra thing I found: When I put all the covers back and ran the
scope for a while, I found that the one screw on the right hand side
is actually quite important.
This creates a thermal connection between where the internal heatsink
is and the outside blue cabinet, so that way the scope's casing helps
in cooling the scope too.

Again, pictures are here:
/g/TekScopes/album?id=64919&p=pcreated,,,20,1,0,0

Un saludo,

Leo
I'm in a 120 volt world and I just had a "Schaffner filter failure" in
a 2230 about a month ago.
What an nasty smell!
That's two this year.
You may be able to change it by just removing the two screws and pulling
it out.
You will need a hefty iron to unsolder it.

-Bert
Hi Bert,

I am in Spain, but based on the SN (204xxx) my 2215A was manufactured in the UK, so it's probably always been in a 220V environment.
Another group user Tom send me some documents about modifications done to the PS, but I got the impression these mod's were based on the early US build SN's, depending on scope model. (01xxxx - ), and not for the later manufactured UK (2xxxxx) and Dutch (7xxxxx) SN's?
I checked my scope for these mod's, and some were standard in the schematic, but funny enough there seem to be many iterations around of the PS design, some mod's are to be done only to 2235's, and some only to 2215 but not to 2215A, etc.

What I did do was:

Replaced the Schaffner line filter to be on the safe side;
Replaced the two (already micro cracked) 2n2 RIFA capacitors in the extra line EMI filter build to the back of the scope;
Kept the capacitor C900, it looks as if that one was replaced before;

Pictures of the removed compnents are here: /g/TekScopes/album?id=64919&p=pcreated,,,20,1,0,0

Thanks for the input,

Leo


Re: 2215A LVPS repair [Solved]

 

On 8/16/2018 9:43 AM, tom jobe wrote:
Hi Leo,
Thanks for the progress report, It's good to see that your 2215A repairs so far have gone well!
In studying your photos, it looks like you have one of those dreaded "Schaffner" brand AC inlet filters (where the AC cord plugs into the back of the scope).
If that is what you have, they have a poor reliability record when used on 240 AC line voltage, if you also have that too.
I'm in a 120 volt AC part of the world and I started changing them after having just one scary "Schaffner" failure.
The Tekscopes message archive should have some "Schaffner" discussions from the past, and a Google search on the term 'Schaffner filter failure' will get you more information too.
You might also consider changing the few X and Y caps in the 2215A, especially C904 which is under the AC inlet filter and behind the main power switch.
C904 is much easier to change while you have the AC inlet filter removed.
The original X and Y capacitors have difficulty with 240 volt mains, so you might want to get the highest rated replacements you can find.
tom jobe...

On 8/16/2018 4:19 AM, satbeginner wrote:
Hi all,

just a final message: it's all done!

Today I received the 12V 60mm brushless fan that was a perfect fit.
I used the half wave rectifier circuit I was able to place the components for on the board, the fan is running nicely without real noise on 9V.
I mean, you can hear it, but only just.
There are plenty perforations all around the casing to let cool air in, so I choose to have it blowing out, so it is sucking the warm air from the inside of the scope and remove it from the scope.
(Like in my 2465's)

One extra thing I found: When I put all the covers back and ran the scope for a while, I found that the one screw on the right hand side is actually quite important.
This creates a thermal connection between where the internal heatsink is and the outside blue cabinet, so that way the scope's casing helps in cooling the scope too.

Again, pictures are here: /g/TekScopes/album?id=64919&p=pcreated,,,20,1,0,0

Un saludo,

Leo
I'm in a 120 volt world and I just had a "Schaffner filter failure" in a 2230 about a month ago.
What an nasty smell!
That's two this year.
You may be able to change it by just removing the two screws and pulling it out.
You will need a hefty iron to unsolder it.

-Bert




Tektronix 465,NO Trigger

 

Greetings All, I recently purchased a 465 on Ebay UK, having now cleaned it up it's looking good, It seems to be working with a good sine wave on both channels from the calibrator, but the only way I can trigger the waveform is by very careful adjustment on the TIME/DIV control and the little red Knob in the center, I don't have a Trigger light ( the lamp in OK) so I presume that I have a Trigger fault. I have checked all the voltages and all are in spec.
Although I am fine when repairing Ham Transceivers I know nothing about Scopes.
Can any kind expert give me some suggestions as to what to test first ??
Thanking you in advance
John
mi0dfg


Re: TG501 20ns 10ns 5ns

 

Hello Kerry,
Just a correction...
While the hypothesis of the ¡Â5 stage misbehaving is still sound and would cause exactly the symptoms you describe, I got wrong (and stated wrongly) that the slower timings (slower than 50ns) were derived directly from the 1MHz reference clock. They are not.
All the timings are obtained by division of the 100MHz PLL generated clock.
Still the theory of the ill ¡Â5 divider remains, as it would - at the same time - make the PLLed frequency to be lower (which would directly impact all the time ranges that are BEFORE the ¡Â5 divider) but, since the division factor would be less, all the ranges slower than 50ns (which are divided AFTER the ¡Â5 divider) would remain correct (80MHz ¡Â 4 = 100MHz ¡Â 5).
Rgrds,
Fabio

On Thu, Aug 16, 2018 at 03:02 PM, Fabio Trevisan wrote:


Hello Kerry,
I don't own a TG501 so I don't really qualify as an expert, but I downloaded
the service manual from the Tekwiki (here: )
and, by just looking at the block diagram, it doesn't seem a simple tweak
(e.g., the turn of a trimpot) to change the time marks of 20ns to 25, the 10ns
to 12.5 and 5n to 6.25.
Still, it's technically possible that only the 20n and faster ranges are off,
since all the slower ranges are derived by direct division of the 1MHz
reference clock, while the faster ranges, rely on a 100MHz clock internally
generated by PLLing the 1MHz reference clock, which is divided by 100 (in
successive divisions, ¡Â5, ¡Â2, ¡Â5 and ¡Â2) and phase compared to the
reference 1MHz clock.
If the error is exact (i.e. if the 20ns is turned into exact 25ns) a good
theory could be that something wrong with the 1st stage of dividing by 5
(U310A, U310B and U315A) is making it divide by 4 instead.
This would make the PLL to generate an 80MHz frequency, instead of 100MHz.
First hypothesis that came to my mind was that a decade divider IC could have
been replaced by a binary divider... But not applicable in this case, as the
¡Â5 is obtained by a discrete combination of 3 ECL flip-flops.
My second hypothesis is that, maybe a false contact, (a bad IC socket comes to
mind) is interrupting some signal in this ¡Â5 logic, causing it to misbehave
and divide by 4.
In face of the known issues that some Texas sockets used by Tektronix on the
equipment of this vintage, it's worth to give it a check.
See this link () for more
information on the dreadful Texas IC socket.

If, however, the error is not exact... than it may just be that the PLL is
failing to generate the 100MHz clock.
In this case, it can be easily diagnosed (even with an uncalibrated scope), by
just displaying simultaneously (on 2 channels), the 1MHz reference clock, at
the same time with any of the clocks generated from the - supposed - 100MHz
clock.
If you trigger in one and the other roll on the screen, or vice-versa, than
it's the PLL that it's not locking.

Hope this helps,

Rgrds,

Fabio




On Thu, Aug 16, 2018 at 01:14 PM, <kburns@...> wrote:


Hello

I'm a newcomer to this group who has fairly recently acquired a SG503,
PG506
and TG501 in a TM5003 module. All of these units are working fine, apart
from the TG501 which shows some odd outputs on the 20ns, 10ns and 5ns
markers. The timing seems to be correct on the slower markers (50ns and
below), but the 20ns, 10ns and 5ns show as 25ns, 12.5ns and 6.25ns
respectively. The 2ns is correct (can't check the 1ns, given the bandwidth
of
my old Tek 2465A). I've checked the markers on a second scope as well.
The
TG501 is a later model (Bo51xxx).

My first thought was that there must be a fault with the 20ns, 10ns and 5ns
outputs - possibly in the dividers - but then wondered if perhaps the
generator had been deliberately modified at some point to give the 25ns,
12.5ns and 6.25ns values. I'm not sure if this mod would be possible with
the TG501? I've had a good look at the manual, but before I start delving
any further I wanted to be sure it is a fault. I have checked the power
supply voltages and ripple and the 1 Mhz oscillator - these are all good.

Any advice on this issue from the experts here would be appreciated.

Thank you

Kerry


Re: 2215A LVPS repair

tom jobe
 

Hi Leo,
I didn't see the photo(s) of your Schaffner AC inlet filter posted, so I put two photos in your photo album of some of the Schaffner filters I have that were taken out of some 22xx scopes.
/g/TekScopes/album?id=64919&p=pcreated,,,20,1,0,0
Some of them have an extra tag on the side with that 'FUY 1651' on them, but others have no extra tag on them. I have more of these Schaffner filters I've removed, but they are all the same.
I think those two X or Y caps in the black filter box on the back wall of the scope are the only other X or Y caps in your 2215A besides C904 on the mainboard.
There was a circuit modification done to that black box where two components get their positions in the circuit switched. I've done this modification several times, but I wonder if it is worth doing.
All of this is described in the instructions for the Mod Kit 050-2242-03 which is available at:
hakanh.com/dl/docs/kitinstructions/050-2242-03.pdf
A few years back I also made a document that relates to all of this, it's not posted anywhere that I know of, so I will send it to you directly.
It turns out that you can not put anything but "image" files in your photo album/
tom jobe...

On 8/16/2018 9:05 AM, satbeginner wrote:
On Thu, Aug 16, 2018 at 03:43 PM, tom jobe wrote:

Hi Leo,
Thanks for the progress report, It's good to see that your 2215A repairs
so far have gone well!
In studying your photos, it looks like you have one of those dreaded
"Schaffner" brand AC inlet filters (where the AC cord plugs into the
back of the scope).
If that is what you have, they have a poor reliability record when used
on 240 AC line voltage, if you also have that too.
I'm in a 120 volt AC part of the world and I started changing them after
having just one scary "Schaffner" failure.
The Tekscopes message archive should have some "Schaffner" discussions
from the past, and a Google search on the term 'Schaffner filter
failure' will get you more information too.
You might also consider changing the few X and Y caps in the 2215A,
especially C904 which is under the AC inlet filter and behind the main
power switch.
C904 is much easier to change while you have the AC inlet filter removed.
The original X and Y capacitors have difficulty with 240 volt mains, so
you might want to get the highest rated replacements you can find.
tom jobe...

On 8/16/2018 4:19 AM, satbeginner wrote:
Hi all,

just a final message: it's all done!

Today I received the 12V 60mm brushless fan that was a perfect fit.
I used the half wave rectifier circuit I was able to place the components
for on the board, the fan is running nicely without real noise on 9V.
I mean, you can hear it, but only just.
There are plenty perforations all around the casing to let cool air in, so I
choose to have it blowing out, so it is sucking the warm air from the inside
of the scope and remove it from the scope.
(Like in my 2465's)

One extra thing I found: When I put all the covers back and ran the scope
for a while, I found that the one screw on the right hand side is actually
quite important.
This creates a thermal connection between where the internal heatsink is and
the outside blue cabinet, so that way the scope's casing helps in cooling the
scope too.
Again, pictures are here:
/g/TekScopes/album?id=64919&p=pcreated,,,20,1,0,0
Un saludo,

Leo


Hi Tom,

As it turns out, it is not yet solved... ;-)

Yes, I opened the scope again, and yes, it's a Schaffner line-filter.
I put a picture of it in the album: /g/TekScopes/album?id=64919

Is this one of the "explosive" types?? If so, I'll will replace it.
Although this scope spend his whole life in The Netherlands,, so maybe it OK?

Further down the line there is this black, plastic box that contains another line-filter.
In there are two 2n2 RIFA capacitors (similar as in the 2465 PS), they already look "cracked", so I will replace them too.

To be continued,

Leo