Re: Quality of FG 502 Waveforms


 

Thanks Jerry;

I have the same result using a variety of cables into a 50 Ohm feedthru termination on several different scopes.

It has two nearly identical rectifier circuits (+20 and -20V supplies) and in comparing the two, it appears that the +20 side rectifier has a diode that is breaking down in both directions. I am wondering if this could cause both the noisy output and the nonadjustable +20V.

Phil...

--- In TekScopes@..., jerry massengale <j_massengale@...> wrote:

Hi phil,

Seeing the noisy signal makes me want to check shields and grounds. maybe a bad coax cable. Do you have the proper load on the output?