Re: Tek 7854 intermittent issues


 

The digital board supply is only accessible if you remove all the logic boards - it hides away at the front of the 'scope to the left (as viewed from the front) of the logic boards.

You can't get at that board while the scope is running, so you have look at the schematics and board layouts to find places to check the -12V and +/-5V that it supplies to the logic boards.

Refer to pp 2-124 and 2-125 for a description of that supply and Schematic <38> for the A7 Auxiliary regulator board.

You really need extenders to work on this beast - Gerber files available - sorry I have NO SPARE edge connectors.

-----Original Message-----
From: [email protected] [mailto:[email protected]] On Behalf Of Nick Corvid
Sent: 05 September 2020 06:36
To: [email protected]
Subject: Re: [TekScopes] Tek 7854 intermittent issues

Alright I've checked all the power supply test points I knew the location of (A12,A22,A17) Except I didn't get that digital supply David mentioned, which I'm not sure where the test points are, how to reach them, or what the specs are for that.

Of the stuff I checked though everything came out relatively normal from what I could measure with my less than great instruments, well all but one thing, TP126 for the +108 volts that read 108.5 volts which is barely in spec but had some funky ripple pictured here on my auxiliary scope: /g/TekScopes/photo/252782/0?p=Created,,,20,2,0,0 . That was the only thing I saw that was out of the ordinary so far.

Join [email protected] to automatically receive all group messages.