Re: Agilent 3458A Troubleshooting
¿ªÔÆÌåÓýHi Congrats Try a look at eevblog---they have done alot of work with HP 3458. Regards Hardy ? Fra: [email protected]
[mailto:[email protected]] P? vegne af Labguy ? I am trying to restore an HP3458A DVM and I need a little
help. |
Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns!
Hey, bad-mood-Dave! Children takes the bad quality chinese crap because of price. I was (still I am, surely) a very poor guy and had to use bad quality tools in my childhood. And suffered the effects on that. When I went "pro" I understood the value of good, quality and expensive tools. My first weller soldering iron I got when I was 15, and I still use it!? But for hobby...not everyone can afford an E8285A. I have one just because I got it for free. Around $2000 here in Brazil, too expensive fot 98% of all us Brazilians I am happy children are buying crap tools. In the past theu couldn't afford anything. Been there... 73 de alexandre pu2sex Enviado do meu Tele-Movel Em qui, 17 de mar de 2022 22:28, Dave McGuire <mcguire@...> escreveu: On 3/17/22 21:16, Lothar baier wrote: |
Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns!
I think I might still have one or two.? Of course they are perfectly safe unless you mistreat them, and that applies to so many things.? I'm always on the lookout for Hg in old thermostats, etc.? Both to keep it out of the landfill, and to add to my collection.? I'm glad I grew up when I did; I'd hate to be a kid these days when everything is so "safe"!? ? -Dave ? It has been strange to watch the pendulum swing over the years in that respect. I have known a time when we still had silver thermometers. ? Zen From: [email protected] <[email protected]> On Behalf Of Radu Bogdan Dicher
Sent: Saturday, March 19, 2022 12:10 AM To: [email protected] Subject: Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns! ? Dave, Eric! (as I've known you as Zen over emails), ? Agewise, I'm probably right between you two, though?I'll wager far closer to Eric than Dave. Likely at least the same generation. I've been doing tubes for a good bit over 15 years, and repeatedly getting to taste the 450V of my amps' B+ hasn't ever deterred me. I started with electronics at about 7, and no one watched me much if at all at the club house when I was etching my boards with acid.? ? That said, I do want to suggest a certain amount of reserve with how we feel about the dangers of industry from certain times. I grew up in the most polluted place in Europe in the '80s (to substantiate, this orange acid rain was coming down most evenings around 6:30PM, so no one would ever be out for a couple of hours around that time, particularly in the fall, due to factory schedule of heavy chemicals waste release through unfiltered tall towers and atmospheric?conditions on this riverbed valley, and hanging clothes to dry was a lottery in terms of whether you'd be picking them black when dry due to other chemical agents being released in the air at other times, and for kms around this factory, up and down that same valley, the landscape was factually lunar as there was no growth on the bare, red, soil) and I can tell you living?with heavy metals around you, and inside you, does have effects on your body (and probably mind). I am saying this based upon my experience, and growth as a young boy. I am very careful with my kid around my bench.? ? Now, all this said, I want to make sure I thank you, Dave, for encouraging me on a certain project a while back (the younger hobbyist tackling a b..ch of a problem). I can't overstate how important this is, for folks from newbies to maybe some of the veterans. Some of these projects are not easy - if they were, no one would ask for help, and this forum would probably have little reason for being. I don't think we tend to have enough regard for encouraging the OP (sweating and swearing at their bench over a perplexing issue, and probably hating their project a bit at the point of asking for input), and making them feel as part of a community. THIS IS HUGE. I almost never capitalize, but here it's called for, and I am.? ? I am stating this to point out Dave's points, demeanor, and moderation on this forum. Fantastic job at this, Dave.? Radu.? ? On Fri, Mar 18, 2022 at 7:44 PM Jim Ford <james.ford@...> wrote:
|
Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns!
Time is also a problem to long in the tank is a guaranteed over etch.
toggle quoted message
Show quoted text
-----Original Message-----
From: [email protected] <[email protected]> On Behalf Of Chuck Harris Sent: Friday, March 18, 2022 11:57 PM To: [email protected] Subject: Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns! Whether or not a board under etches traces, or over etches traces is entirely dependent on the resist used. Insulating resists don't create any fields in the tank, so they tend to over etch the traces: \^^^^^^^^/ . Metallic resists (gold, solder) do create electric fields that electrolytically deflect the etching action away from the resist, so they tend to under etch the traces: /^^^^^^^^^\ . (Or is it the other way around? It has been so long ago.) -Chuck Harris On Sat, 19 Mar 2022 00:52:13 +0000 "Lothar baier" <Lothar@...> wrote: Eventually dad bought a etch tank from isel , it used a different |
Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns!
¿ªÔÆÌåÓýRadu, ????????????? Totally agree. I have spent some time in what we will call heavy industry. It is one thing to know that a Tek 576 can light you up with 1500 Vdc and might kill you. It is QUITE another to stand next to a peace of equipment that will DEFFINATLY kill you and not even notice that you are in the way. Safety is key in any lab or industrial environment. Respect and caution is needed, abject terror can also be harmful. We don¡¯t need a full flame retardant hazmat suit to deal with an isopropyl alcohol spill. Though HR might want me to use one. ?I have run in to quite a few that won¡¯t take a failed power supply apart and even look at it ¡°because it plugs in to the wall¡± one of my goals is to empower people especially just starting out. Those first few successes are critical. But if one of the first projects is a CRO scope it is also good for some one to say hey don¡¯t touch the fat red wire. Weird stuff like insulation doesn¡¯t work so well at 21,000 Vdc ? It has been strange to watch the pendulum swing over the years in that respect. I have known a time when we still had silver thermometers. ? Zen From: [email protected] <[email protected]> On Behalf Of Radu Bogdan Dicher
Sent: Saturday, March 19, 2022 12:10 AM To: [email protected] Subject: Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns! ? Dave, Eric! (as I've known you as Zen over emails), ? Agewise, I'm probably right between you two, though?I'll wager far closer to Eric than Dave. Likely at least the same generation. I've been doing tubes for a good bit over 15 years, and repeatedly getting to taste the 450V of my amps' B+ hasn't ever deterred me. I started with electronics at about 7, and no one watched me much if at all at the club house when I was etching my boards with acid.? ? That said, I do want to suggest a certain amount of reserve with how we feel about the dangers of industry from certain times. I grew up in the most polluted place in Europe in the '80s (to substantiate, this orange acid rain was coming down most evenings around 6:30PM, so no one would ever be out for a couple of hours around that time, particularly in the fall, due to factory schedule of heavy chemicals waste release through unfiltered tall towers and atmospheric?conditions on this riverbed valley, and hanging clothes to dry was a lottery in terms of whether you'd be picking them black when dry due to other chemical agents being released in the air at other times, and for kms around this factory, up and down that same valley, the landscape was factually lunar as there was no growth on the bare, red, soil) and I can tell you living?with heavy metals around you, and inside you, does have effects on your body (and probably mind). I am saying this based upon my experience, and growth as a young boy. I am very careful with my kid around my bench.? ? Now, all this said, I want to make sure I thank you, Dave, for encouraging me on a certain project a while back (the younger hobbyist tackling a b..ch of a problem). I can't overstate how important this is, for folks from newbies to maybe some of the veterans. Some of these projects are not easy - if they were, no one would ask for help, and this forum would probably have little reason for being. I don't think we tend to have enough regard for encouraging the OP (sweating and swearing at their bench over a perplexing issue, and probably hating their project a bit at the point of asking for input), and making them feel as part of a community. THIS IS HUGE. I almost never capitalize, but here it's called for, and I am.? ? I am stating this to point out Dave's points, demeanor, and moderation on this forum. Fantastic job at this, Dave.? Radu.? ? On Fri, Mar 18, 2022 at 7:44 PM Jim Ford <james.ford@...> wrote:
|
Re: HP83595A/B (HP 8350) ROM firmware required
Thanks!
toggle quoted message
Show quoted text
-----Message d'origine-----
De : [email protected] <[email protected]> De la part de Dave McGuire Envoy¨¦ : 18 mars 2022 22:09 ? : [email protected] Objet : Re: [HP-Agilent-Keysight-equipment] HP83595A/B (HP 8350) ROM firmware required If you don't need version 7 specifically, version 9 for 83595 is available from the KO4BB archive: -Dave On 3/18/22 22:02, Yves Tardif wrote: Hi, I have 8350B ver7 with 83595A with version 3, have you found -- Dave McGuire, AK4HZ New Kensington, PA |
Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns!
Dave, Eric! (as I've known you as Zen over emails), Agewise, I'm probably right between you two, though?I'll wager far closer to Eric than Dave. Likely at least the same generation. I've been doing tubes for a good bit over 15 years, and repeatedly getting to taste the 450V of my amps' B+ hasn't ever deterred me. I started with electronics at about 7, and no one watched me much if at all at the club house when I was etching my boards with acid.? That said, I do want to suggest a certain amount of reserve with how we feel about the dangers of industry from certain times. I grew up in the most polluted place in Europe in the '80s (to substantiate, this orange acid rain was coming down most evenings around 6:30PM, so no one would ever be out for a couple of hours around that time, particularly in the fall, due to factory schedule of heavy chemicals waste release through unfiltered tall towers and atmospheric?conditions on this riverbed valley, and hanging clothes to dry was a lottery in terms of whether you'd be picking them black when dry due to other chemical agents being released in the air at other times, and for kms around this factory, up and down that same valley, the landscape was factually lunar as there was no growth on the bare, red, soil) and I can tell you living?with heavy metals around you, and inside you, does have effects on your body (and probably mind). I am saying this based upon my experience, and growth as a young boy. I am very careful with my kid around my bench.? Now, all this said, I want to make sure I thank you, Dave, for encouraging me on a certain project a while back (the younger hobbyist tackling a b..ch of a problem). I can't overstate how important this is, for folks from newbies to maybe some of the veterans. Some of these projects are not easy - if they were, no one would ask for help, and this forum would probably have little reason for being. I don't think we tend to have enough regard for encouraging the OP (sweating and swearing at their bench over a perplexing issue, and probably hating their project a bit at the point of asking for input), and making them feel as part of a community. THIS IS HUGE. I almost never capitalize, but here it's called for, and I am.? I am stating this to point out Dave's points, demeanor, and moderation on this forum. Fantastic job at this, Dave.? Radu.? On Fri, Mar 18, 2022 at 7:44 PM Jim Ford <james.ford@...> wrote:
|
Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns!
Lothar baier
So the challenge you run into when building replacement modules ( and I am entertaining the idea right now for the power amplifier MIC in the 8753 sources ) is that first of most microwave circuits use ceramic and are wire bonded , although substrates exist that offer similar dielectric properties than Al2O3 wire bonding on those substrates is not trivial !
toggle quoted message
Show quoted text
The bigger challenge though is to figure out the specs of the original module , I been thinking about using a bad 8753 source and rigging it up with some test points to measure power and DC voltages going into the power amp and the loss from output to the step attenuator but right now I haven¡¯t found the motivation yet -----Original Message-----
From: [email protected] <[email protected]> On Behalf Of Dave McGuire via groups.io Sent: Friday, March 18, 2022 9:22 PM To: [email protected] Subject: Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns! On 3/18/22 22:19, Lothar baier wrote: If you do edge coupled microstrip filters or directional couplers thanYes. There's a great deal of activity in building replacement boards and modules for instruments like ours (trying to bring this further on-topic), so if the results there are adequate, this does open up some possibilities. One thing that causes trouble with the more affordable board houses is that sometimes they change substrate compositions, so the published dielectric constant isn't always something you can count on. -Dave -- Dave McGuire, AK4HZ New Kensington, PA |
Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns!
Lothar baier
Back when I worked at UT as a research engineer we had a laser cutting system for MMIC Dies , it was a interesting device but required a lot of skill and experience to use , when you looked through the scope it looked like carpet bombing ?
toggle quoted message
Show quoted text
-----Original Message-----
From: [email protected] <[email protected]> On Behalf Of Bryce Schroeder via groups.io Sent: Friday, March 18, 2022 9:22 PM To: [email protected] Subject: Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns! I've looked at the laser'd traces from above a lot, tested their isolation with a megger, etc and found them all satisfactory, but I don't know how the roughness is side-on. It would be interesting to investigate. The laser is fundamentally a pulsed device, but the pulse rate is in the 10s of kHz (up to maybe a MHz depending on the particular kind of laser) and cutting through even ordinary half oz copper is a lot of individual passes. Definitely something to investigate. On Fri, 18 Mar 2022 at 22:19, Lothar baier <Lothar@...> wrote:
|
Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns!
Lothar baier
With FR-4 you definatly have issues with ER lot variations this is why you use materials like rogers 4350B
toggle quoted message
Show quoted text
-----Original Message-----
From: [email protected] <[email protected]> On Behalf Of Dave McGuire via groups.io Sent: Friday, March 18, 2022 9:22 PM To: [email protected] Subject: Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns! On 3/18/22 22:19, Lothar baier wrote: If you do edge coupled microstrip filters or directional couplers thanYes. There's a great deal of activity in building replacement boards and modules for instruments like ours (trying to bring this further on-topic), so if the results there are adequate, this does open up some possibilities. One thing that causes trouble with the more affordable board houses is that sometimes they change substrate compositions, so the published dielectric constant isn't always something you can count on. -Dave -- Dave McGuire, AK4HZ New Kensington, PA |
Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns!
Whether or not a board under etches traces, or over etches traces
is entirely dependent on the resist used. Insulating resists don't create any fields in the tank, so they tend to over etch the traces: \^^^^^^^^/ . Metallic resists (gold, solder) do create electric fields that electrolytically deflect the etching action away from the resist, so they tend to under etch the traces: /^^^^^^^^^\ . (Or is it the other way around? It has been so long ago.) -Chuck Harris On Sat, 19 Mar 2022 00:52:13 +0000 "Lothar baier" <Lothar@...> wrote: Eventually dad bought a etch tank from isel , it used a different |
Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns!
Lothar baier
¿ªÔÆÌåÓýA few words on FR-4 and RF boards ,? FR-4 generally ¡°works¡±? ok for RF boards up to 2GHz but if you use coupled microstrip filters or directional couplers you are going to have the time of your life getting repeatable results ! First of the dielectric constant of FR-4 is not tightly controlled and can vary from batch to batch , dielectric loss is also higher compared to microwave materials ! Generally for most applications rogers 4350 or 4003 are popular ,? the cost is competitive , the boards are fairly ridged comparable to FR4 and most FR4 lines can process them with no problems making it easy for board manufacturers . For mm wave applications or if extremely low losses are required 5880 or 5880 are commonly used but those substrates require special processing especially for plated through holes and the materials are expensive ? From: [email protected] <[email protected]>
On Behalf Of Jim Ford via groups.io
Sent: Friday, March 18, 2022 9:44 PM To: [email protected] Subject: Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns! ? Definitely true for garden-variety FR-4 board material.? For serious microwave work, say >6GHz, you'd want a more stable material, like those made by Roger's, Isola, etc. ? Jim Ford ? ? ? Sent from my T-Mobile 4G LTE Device ? ? -------- Original message -------- From: Dave McGuire <mcguire@...> Date: 3/18/22 7:22 PM (GMT-08:00) Subject: Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns! ? On 3/18/22 22:19, Lothar baier wrote: |
Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns!
Back in 1972, when I worked for DEI, I did photo resist
application on circuit boards, and it is trivial. We had a stainless steel rectangular tank, big enough so you could vertically dip the biggest card we did, but small enough so that it held a minimal amount of chemicals. The tank was bent out of sheet metal in our shop, and had a little S.S. lid, like a gift box, that covered the tank when we weren't using it. We cleaned the copper until it was shiny, and did a final clean with tri-chlor, and then simply dipped it in the KPR and extracted it, letting it drip dry in a dark space. It's that easy. I am sure that someone still makes KPR, though I have no idea who. Exposure was done as a contact print, with a small carbon arc lamp as the source. It was developed by dipping the exposed board in tri-chlor, where the unexposed KPR would wash away. -Chuck Harris On Sat, 19 Mar 2022 01:04:22 +0000 "Lothar baier" <Lothar@...> wrote: I stopped doing it years ago mainly because the materials I mostly |
Re: Agilent 3458A Troubleshooting
tmillermdems
¿ªÔÆÌåÓýYou can get the clip from Artekmedia. The first step is always to check all the low voltage power supply voltages. The calibration on the 3458a just needs a precision 10 volt reference and an accurate 10 k resistor. It is a wondeful instrument but repair and cal Keysight is a bit painful. It is still in production. Regards and good luck.
On 3/18/2022 10:30 PM, Labguy wrote:
I am trying to restore an HP3458A DVM and I need a little help. |
Re: Agilent 3458A Troubleshooting
On a quick note - just wanted to congratulate you for the score - assuming you didn't pay an arm and a leg for it (as they tend to go for, last I checked).? Good luck with the repair! I'm certain this group will help to reach a positive resolution. Radu.? On Fri, Mar 18, 2022 at 7:32 PM Labguy <georgg@...> wrote: I am trying to restore an HP3458A DVM and I need a little help. |
Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns!
¿ªÔÆÌåÓýDefinitely true for garden-variety FR-4 board material.? For serious microwave work, say >6GHz, you'd want a more stable material, like those made by Roger's, Isola, etc. Jim Ford Sent from my T-Mobile 4G LTE Device -------- Original message -------- From: Dave McGuire <mcguire@...> Date: 3/18/22 7:22 PM (GMT-08:00) Subject: Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns! > If you do edge coupled microstrip filters or directional couplers than the accuracy of the gaps is very important but another aspect that has to be considered is the roughness of the edges ?? Yes.? There's a great deal of activity in building replacement boards and modules for instruments like ours (trying to bring this further on-topic), so if the results there are adequate, this does open up some possibilities.? One thing that causes trouble with the more affordable board houses is that sometimes they change substrate compositions, so the published dielectric constant isn't always something you can count on. ??????????? -Dave -- Dave McGuire, AK4HZ New Kensington, PA |
Agilent 3458A Troubleshooting
I am trying to restore an HP3458A DVM and I need a little help.
I acquired this unit on ePay some time ago and I finally found some time to check it out. The outside and inside are both neat and clean. The build date is 1988, but someone has maintained over the years, because it has the latest firmware ROMs fitted (Rev 9). The Dallas NV RAMs are 2012, so they are nearing end of life. It is fitted with Option 002? Improved DC reference stability 4 ppm/year(!). The Cal sticker says it was last calibrated in 2013. A visual check of the PCBs revealed nothing obvious, so after setting the mains switch to 230V I decided to power it up. I was greeted with a sizzle and an acrid smell. I have been there before with other old test gear - the Schaffner mains filter kacked itself. So I ordered a new one and replaced it. When I powered the unit on again, I was greeted with a short BEEP and the first rectangular block of the green electroluminescent display flashes ON and OFF. That's as far as it gets. I have searched the Agilent manuals, all the info I could find on the internet and watched some excellent Youtube clips on repairing and calibrating 3458A's. I cannot find any troubleshooting reference for a flashing single character block on the display. I am thinking there may be something wrong with the firmware code in one of the EPROMs that is causing it to hang while booting. I have an EPROM reader/burner, but I am reluctant to go there until I am sure I have to. I would be most grateful for any advice anyone may care to offer. Cheers, George |
Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns!
Lothar baier
Laser toner wont work for microwave circuits for a variety of reasons¡..
toggle quoted message
Show quoted text
The way i did my boards was to plot out a layout magnified 10:1 , then dad took the thing to work and used a repro camera to scale it and create a negative film , the film was put on a photo resist coated board , exposed and developed. Once developed the bare copper showed where the traces would be , the board then went into a plating tank and was gold or silver plated , once the plating was done board was stripped and etched. On Mar 18, 2022, at 21:19, Lothar baier <Lothar@...> wrote: |
Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns!
Sure, Dave, I'll send you an email off list to request the address. Thanks!
toggle quoted message
Show quoted text
On Fri, 18 Mar 2022 at 22:20, Dave McGuire <mcguire@...> wrote:
|
Re: BS, was Re: [HP-Agilent-Keysight-equipment] Spread the word -- Swap Meet Returns!
I've looked at the laser'd traces from above a lot, tested their
toggle quoted message
Show quoted text
isolation with a megger, etc and found them all satisfactory, but I don't know how the roughness is side-on. It would be interesting to investigate. The laser is fundamentally a pulsed device, but the pulse rate is in the 10s of kHz (up to maybe a MHz depending on the particular kind of laser) and cutting through even ordinary half oz copper is a lot of individual passes. Definitely something to investigate. On Fri, 18 Mar 2022 at 22:19, Lothar baier <Lothar@...> wrote:
|