Re: Agilent E8357A Network Analyzer Phase Lock Lost Error
开云体育That is pretty bad. An ovenized crystal oscillator should be able to do at least 10^-7. – Mike ? Mike B. Feher, N4FS 89 Arnold Blvd. Howell NJ 07731 908-902-3831 ? From: [email protected] <[email protected]> On Behalf Of Johnny10 via groups.io
Sent: Sunday, May 25, 2025 1:04 PM To: [email protected] Subject: Re: [HP-Agilent-Keysight-equipment] Agilent E8357A Network Analyzer Phase Lock Lost Error ? Actually measuring again with a counter I come up with 10.000447 Not to say my equipment is perfectly accurate. ? |
Re: Checking an 8402B Calibrator
It might be simpler to put a 100M ohm resistor in series with the high input on the 34401A in the High Z mode. Your 17.8V reading will now be 8.9V well inside the 10V limit for the 10V, 100M ohm impedance range. If required you can calculate the exact division ratio and correct your reading accordingly.
Don Bitters |
Re: HP8510C "system bus address error", but then it works
开云体育?Marko, The system may have been set up to use two sources and it’s looking for the second source. Try: Instrument State <Recall><More><Factory Preset> Regards, Steve On May 25, 2025, at 2:14?PM, Marko Cebokli via groups.io <s57uuu@...> wrote:
|
HP8561E Jittery Display
Hi, I've some HP specans just obtained as BER items from a local space communication site.
The first one I'm looking at is an HP8561E, it powers on OK and the display is bright but the text is jittery and almost unreadable.
I'm guessing "check out the PSU, Paul" is the way to go but does anyone have any insight on this model/fault they might like to share?
I'm also looking for component level schematics etc.
?
The other items I have are all HP859x series.
?
Any info welcome thanks
Paul
?
? |
HP8510C "system bus address error", but then it works
Got my hands on a 50GHz 8510C with test set and source, everything mounted in a mobile rack with a pullout table.
When I switch it on, it says "caution: system bus address error", but them it proceeds to function OK.? Both the test set and source respond to commands. I have checked the addresses, and they are set as specified in the manual, both on the instruments and in the HP-IB setup menu.
?
Not much of a problem, since I can use the analyzer, just a bit annoying. Any idea how to fix this?
?
Marko Cebokli |
Re: HP 8902A Measuring Receiver
Hey there Raymond,
?
I've done a little futzing with the power supply and the input amp on my 8902, so I'm happy to try and help where I can.
?
I think you might get more traction on the group if you describe the problem you have and see who might have ideas about how to address.
?
TonyG |
Re: ftp.agilent.com tar-file on archive.org
I have been downloading files with "wget" for many years, but never used the "-c" option (I guess as long as one just downloads "regular" size files, it doesn't matter that much).
?
It worked like a charm! Thank you for the tip.
?
Best,
Wolfgang
KI7PFX
?
On Thu, May 15, 2025 at 05:23 PM, Wolfgang Schraml wrote:
|
Re: Checking an 8402B Calibrator
Tony,
If you have or can borrow one, the Fluke 8840A does High-Z up to 20Vdc.
--John Gord
?
On Sat, May 24, 2025 at 09:03 PM, Tony Goodhew wrote:
|
Re: Checking an 8402B Calibrator
Thanks, Jim & Charles - Appreciate it.
?
Where I find myself is that the 8402B manual says to look at a voltage that is around 17.8 Vdc above the -30V rail (Section 5-7 in the manual) - This pushes my 34401A into the 100V range, which means it is only 10Mohms input impedance (there are two measurements that are here, otherwise I can just use the 10G setting and do the other ones) - This is why I think I need to implement a voltage follower with an OpAmp to get the >100M input impedance.
?
Having never done this before, I'd love guidance from anyone who has done this or worked with the 8402 - Coming back to the reason I'm asking is that I have a 431C which was broken and I think I've fixed it but I sort of need the 8402 to confirm that I have done that (Old lady swallowing fly and all).
?
Thanks,
?
TonyG |
Re: Checking an 8402B Calibrator
开云体育Try using the 34401A in hi-Z (high impedance) mode, Tony. ?I had to do that once about 10 or 15 years ago when measuring voltage in a low current circuit. ?Got weird results until I switched it to hi-Z mode. ? HTH.Jim Ford Laguna Hills, California, USA? On May 24, 2025, at 6:17?PM, Tony Goodhew via groups.io <tony_goodhew@...> wrote:
|
Checking an 8402B Calibrator
Hey All,
?
I'm looking at an 8402B calibrator, and the manual says that the measurements need to be done with a floating DMM that has a greater than 100Mohm input impedance.
?
The manual says to use an HP 740B - I don't have one of these, so I'm thinking of using an HP 34401A with an OpAmp Voltage Follower to buffer between the 8402 and the 34401.
?
Has anyone done this before and could provide any insights?
?
Thanks,
TonyG |
Agilent E8357A Network Analyzer Phase Lock Lost Error
I have been troubleshooting my Network Analyzer using the E8356A, E8357A , E8358A Service Guide June 2004. This 8357 has an upgraded 700 MHz CPU.
After reading and Option 015.
My problem is Phase Lock Lost error in a number of bands. All would seem to relate to the A8 Fractional-N Synthesizer Board Assembly.
Fortunately I could partially run the Serial Bus Test where I found wildly out of spec voltages on the board.
The Serial test node numbering on my application do not coincide with the manual Nodes but the description places me in similar territory as the Manuals Node descriptions.
The test has a minimum and maximum range for the Node as I will describe.
??????????????????????????????????????????? Min????????????? Result?????????? Max
+5Vdc Supply Voltage???? 4.28.?????????????? 5.0?????????????? 5.78.???? OK
Band ALC Level???????????????? -8.27?????????? 14.0169????????? 1.0??????? Bad
VC01 Tune Voltage???????? ?? 6.3???????????? -1.345 ?????????? 14.3???? Bad
VCO Compensation ? ?? ?? 2.07??????????? 13.748??????????? 4.07????? Bad
????????
?
Node descriptions A11 Phase Lock Board. ( the signals to the A11 board 30kHz block come from the A8 Fractional-N board.
30kHz Phase Detector???? -0.33.???????? -5.088 ? ? ? ? ? 0.33. ? ? Bad?
30kHz Phase Detector Integrator
?????????????????????????????????????????? -2.86?????????? -13.827?????????? 0.44.???? Bad
?
10MHz Frequency.? 10.000000
5Mhz Frequency?????? 4.999756
?
A10 Phase Lock Test??
Network Analyzer Setting?? W39? From A10J5
Band 0???????????? Setting 9.9MHz???????????? Result should be? ??? 9.9MHz ??????????????????? Actual Measurement ??????? 9.87??????? (Very weak signal)
Band 1 - 6.?????? Setting. 1 GHz?????????????? Result should be ?? 1.04166MHZ ??????????? Actual Measurement ?????? 1.041616MHz
?
A8 Fractional -N
Band 0 Output????? Setting 5MHz.?????? Result Should be 120.16 MHz.??? Actual Measurement?????????? No distinct signal
???????
I have added pics from CW Testing showing the frequencies and Bands affected.
And the 20MHz - 550Mhz segment.
?
My questions: 1. Would you agree with A8 Fractional board being bad?
???????????????????????? 2. Are there A8 Boards available in the < = 500 dollar range.
?
? |
Re: Joining group