John and others following this,
The AZ(auto zero function) and Manual/Auto? functions are two different animals.

The AZ function shorts the input to the DC conditioning amplifier, U101, before each measurement of the customer's input to determine any residual offset of the input due to thermal drift and subtract that offset value from the input being measured.? ?So in principle if the input offsets drift over changes of time and temperature with a shorted input, the display will continue to read ZERO.?

Internal switches inside the input hybrid U102 are called? ? ?MZ -measure ZERO? and? MC -measure CUSTOMER (input)

With Auto Zero ON,? two readings are made for each measurement registered on the display:
[1]:? measure the offsets:? ?(MZ switch ON,? ?MC-measure customer switch OFF)? ? ? ? ? ? ? ?

[2]:? ?measure the customer input:? ?(MC switch ON, MZ switch OFF)? ? ? ? ? ??

[3}:? subtract [1]? from [2] and display the value

If the AZ function is turned OFF, the input remains static with the input constantly applied to the input of U101.? The reading rate in this condition is effectively doubled because the MZ and MC switches are no longer being toggled and all measurements are a measure customer with no new offsets being measured.? The ZERO from the last time AZ was active is subtracted from the customer input until AZ is turned on again.? So if you have AZ turned OFF, it is a good practice to turn it on periodically to get a new ZERO reading if the environment is unstable that may cause the input offsets to drift.

Manual range and Auto ranging is independent of the AZ function.? ?The meter can stay fixed on a given range or auto range with or without the AZ function turned ON.
Hope this helps you,

George? ? ? ? March 18, 2019? ? 11:02 am


Re: E8357A upgrade questions

 

Caesar

I am interested in getting the CPU and hard drive. Sent a PM.

1. Is there anything special I need to do for the install? I thought I read somewhere that my front panel needs to be changed? I assume the Bios on the new CPU will be OK.

2. Can I easily revert back if I have an issue??

Tom


Re: 5385A error 52

 

In a few day , I will able to check the hp5386 with the Nationa GPIB PCMCIA card . I suppose the hp5386 is very similar to your counter.
Im not shure to understand very well the problem : you must wait the time out to get response from the counter ?.
If that is your problem / question , there is something not working properly or the setup is not ok , like the string terminator for example.
Best regards
Maurizio


Re: HP 83592 - Low output power on low bands

Pete Manfre
 

I might have a complete spare for a 592 that if I remember has level issues.? ?I will have to look if you are interested.?

Pete wa2odo

On Mon, Mar 18, 2019, 9:34 AM ON7MA <on7ma@...> wrote:
Hello
does anybody have a spare part or old HP83592 A/B/C for sale ?
It seems that the RF switch in the End Module HP SYTF ( Yig Filter ) is broken
The part is 5086-7404
I dont think it is repairable
JP
The lower signal does not come through.. control tension is present...


HP 83592 - Low output power on low bands

 

Hello
does anybody have a spare part or old HP83592 A/B/C for sale ?
It seems that the RF switch in the End Module HP SYTF ( Yig Filter ) is broken
The part is 5086-7404
I dont think it is repairable
JP
The lower signal does not come through.. control tension is present...


Re: HP 5335A power supply check

 

Hi Dan,

Considering that your serial number is 2510A, you need the correct manual. I posted an updated manual on the KO4BB site, so feel free to get it there. The change in voltage on the power supply rail was called out in one of the addenda also posted with the manual).

1) At serial number 2336A and above, the +15.7 V specification changed to +15.000 V. So, even though the 5335A is very forgiving, I would align the counter supply to that +15.000 V voltage.

2) IIRC, the procedure you mention calls for the interconnect cable to be removed, and for pins 1 and 16 on the A1 power supply board be shorted in order to bring up the supply (please carefully check my recollection). The power supply will not power on with the cable removed unless these two pins are shorted together.

I work on these units quite often, so please let me know if you have questions, and I will try to answer them. Congrats on acquiring a great instrument!

Regards,

Joe, KN5U


Re: HP 5335A power supply check

 

Dan,

There was an addendum that change the power supply voltage from +15.700 to +15.000 VDC. You may download an updated manual from the KO4BB Manuals website along with all updates and addenda.

In order to power up the supplies with the interconnect cable removed, you must install a temporary jumper across pins 1 and 16 (please verify my recollection).

regards,

Joe, KN5U


Re: 8662A output oddity

 

On Mon, Mar 18, 2019 at 01:31 AM, jfphp wrote:
There is something wrong with the settings of the SA : VF is too narrow compared with the BW filter. Try with auto for all settings.
All settings on the SA were set to auto. This is the "full band" view of the 8569B for the 0.1-1.7GHz band.?

The problem is not with the SA, the problem I am asking about is in the 8662A: the 8662 generates tremendous broadband noise as it sweeps across 1024.8 MHz as shown in the pictures. At *all* other frequencies it generates a very pure tone as expected.

73, Jim N8ECI


Re: 8662A output oddity

 

Hi, there should be a lit star in the upper right corner for settings which not give a calibrated display, right?
G?ran


Re: 8662A output oddity

 

There is something wrong with the settings of the SA : VF is too narrow compared with the BW filter. Try with auto for all settings.
--------------------------------------------

On Sun, 3/17/19, morestainless via Groups.Io <lancelotburt@...> wrote:

Subject: Re: [HP-Agilent-Keysight-equipment] 8662A output oddity
To: "[email protected]" <[email protected]>
Date: Sunday, March 17, 2019, 11:41 PM

Congratulations on your
fine acquisition.? I just got mine going.? Got some
voltage and calibration adjustments to do?





On Sun, Mar 17, 2019 at 3:32 PM,
hpagilent@...<hpagilent@...>
wrote: Hi, I recently acquired my first 8662A
and while looking at its output on a spectrum analyzer I
notice that there is a strange behavior that takes place in
an approximate 2MHz span centered on 1024.8 MHz. At 1024.8
the noise floor increases by about 20dB and when tuning on
either side of 1024.8 the noise floor decreases until it is
completely back to normal at +/- 1 MHz on either side. I
have included two pictures - one showing normal behavior and
the other with the output tuned to 1024.8. This does not
seem to happen at any other frequency in the unit's
range. Has anyone else seen this? Thank you! 73, Jim
N8ECI


Re: I'm officially lost now with this 8753ES

 

Thank you for all that info, it will be helpful.

Regarding my problem, I am still not sure where it is located.? It could either be in the phase lock board or the source assembly.? I will know for sure when I get the replacement phase lock board.? The YTO tuning voltages are enough off from what the service guide says that it could go either way.

With a signal generator as external source, at 1 MHz the unit locks from roughly 500 kHz to 4 MHz so this means the pretune must get to that range (there's the 3.8 MHz osc to work with too).? 200 MHz high is way off.

It might be interesting to feed the YTO with a power supply and see what currents are required at different target frequencies and compare to what the default pretune is in my unit.

To be continued...

Peter

On 3/17/2019 8:25 PM, pianovt via Groups.Io wrote:
On Sun, Mar 17, 2019 at 10:08 AM, Peter Gottlieb wrote:

Regarding the A3 source PCB, I see that it has two 12 bit DACs and a quad
8 bit DAC.? It also has OC drivers for the switchable attenuator that is
an option in the 8753ET.? Does anyone know what all those DACs are used
for?? The coild drive for the YTO comes from the A11 board, so all that's
left are the FM coil on the YTO and the tune input for the 3.8 GHz cavity
oscillator.? What am I missing?

The DACs are part of the leveling loop. In addition to setting the power level, there are various breakpoints and offset adjustments for the level detector and logamp, a slope adjustment, etc. On the ES (and maybe also D), the ALC range was expanded which made this loop very difficult to control.

Also on this board is the bias and temperature compensated varactor drive circuit for the 3.8GHz oscillator. Then, there are also power supply filters for those microcircuits underneath. As far as I remember, the main coil drive merely passes through this board. For Opt. 006 analyzers, there is all the drive circuitry for switching bands. The 8753D/ES analyzers with internal test sets all have step attenuators, so those are also driven from this board.

This board has given me problems more than once. It's a multi-layer board, so tracing it out could take a life time. I have had open vias on it, one board had a cracked pad where that huge power resistor solders in. The design is mecahnically fragile there, but production instructions required bending the resistor after it's soldered into place. I think that may have caused problems on my unit. I also had a leveling loop which oscillated.

But, I don't think any of this is related to your YTO tuning problem. Keep in mind that the reference channel is derived from the last thick film microcircuit. If you have a bad elastomer contact between the microcircuit and the SMA connector, it can cause problems. I can't however explain why the tuning is off only below 1 GHz RF out.

Vladan


Re: Parting out hp85662A display unit

 

Thanks for the interest y¡¯all. ?They went as a lot very soon after the post. ?If anything goes awry in the transaction, I¡¯ll let everyone know. ?At least someone go a good deal and they¡¯ll be put to use. ?Correct on the A4A2&3; same for the A4A4 bandwidth boards.


HP 5335A power supply check

Dan Nessett
 

Hello,

I bought a used HP 5335A and am checking it for operational correctness. I ran the super check function and all power supplies were shown to be within bounds. However, on page 5-2 of the "Temporary Operating and Service Manual" (manual part 05335-90005), the power supply adjustment procedure identifies two power supplies to check: +15.70 and +3.1. These are adjusted by potentiometers on the power supply board. When I checked the 3.1V supply, it was a little low and I was able to bring it back to 3.1V. However, the 15.70V power supply showed 15V when measured at the cathode of A1CR2 (per the note in 5-14/e). When attempting to bring this power supply voltage up to 15.7V, I could only bring it to 15.44V using the identified pot.

I have the following questions:

1) since the super check is passing, should I worry about the 15.70V power supply? The instrument seems to work (I have replaced the HP10811A that came with it in order to get it to properly measure some oscillators I used as DUTs). The fact that the operating and service manual is titled "temporary" suggests there may have been an update, which may have changed the 15.7V power supply to a 15V power supply.

2) When I attempted to diagnose the 15.7V power supply according to the instructions on page 8-53, the procedure failed. In particular, when I disconnected the Power Supply board (A1) from the Amplifier Support board (A2), I could get no voltage whatsoever from any power supply. The instrument wouldn't power on. So, I am suspicious about these instructions and wonder if they are valid for my instrument. The serial number of my HP5335A is 2510A07784, which is above the serial number specified by the manual (prefixes 2024A and below), so I obtained the manual update changes supplement that covers my serial number (05335-90021 - April 25, 1985). However, there was no mention of a change to the 15.70V power supply value. My question: Am I missing a manual or supplement that would explain the adjustment procedure for the 15.70V power supply?

I would be grateful for any help on this problem.


Re: I'm officially lost now with this 8753ES

 

On Sun, Mar 17, 2019 at 10:08 AM, Peter Gottlieb wrote:
Regarding the A3 source PCB, I see that it has two 12 bit DACs and a quad 8 bit DAC.? It also has OC drivers for the switchable attenuator that is an option in the 8753ET.? Does anyone know what all those DACs are used for?? The coild drive for the YTO comes from the A11 board, so all that's left are the FM coil on the YTO and the tune input for the 3.8 GHz cavity oscillator.? What am I missing?
The DACs are part of the leveling loop. In addition to setting the power level, there are various breakpoints and offset adjustments for the level detector and logamp, a slope adjustment, etc. On the ES (and maybe also D), the ALC range was expanded which made this loop very difficult to control.

Also on this board is the bias and temperature compensated varactor drive circuit for the 3.8GHz oscillator. Then, there are also power supply filters for those microcircuits underneath. As far as I remember, the main coil drive merely passes through this board. For Opt. 006 analyzers, there is all the drive circuitry for switching bands. The 8753D/ES analyzers with internal test sets all have step attenuators, so those are also driven from this board.

This board has given me problems more than once. It's a multi-layer board, so tracing it out could take a life time. I have had open vias on it, one board had a cracked pad where that huge power resistor solders in. The design is mecahnically fragile there, but production instructions required bending the resistor after it's soldered into place. I think that may have caused problems on my unit. I also had a leveling loop which oscillated.

But, I don't think any of this is related to your YTO tuning problem. Keep in mind that the reference channel is derived from the last thick film microcircuit. If you have a bad elastomer contact between the microcircuit and the SMA connector, it can cause problems. I can't however explain why the tuning is off only below 1 GHz RF out.

Vladan


Re: HP5335A bottom cover

Dan Nessett
 

Thanks to everyone for their comments. Now, its on to another issue in another thread.


Re: Parting out hp85662A display unit

 

I was going to say A4A2 and A4A3 should be sold as a pair - but it's a moot point if you take all the A4 cards.


On Sun, Mar 17, 2019 at 2:48 PM Stephen Hanselman <kc4sw.io@...> wrote:
Jeff,

Was the display for a ¡°B¡±?? If so I would like all of the A4 cards please

Regards,

?

Stephen Hanselman

Datagate Systems, LLC

3107 North Deer Run Road #24

Carson City, Nevada, 89701

(775) 882-5117?office

(775) 720-6020?mobile

s.hanselman@...

a Service Disabled, Veteran Owned Small Business

DISCLAIMER:

This e-mail and any attachments are intended only for use by the addressee(s) named herein and may contain legally privileged and/or proprietary information. If you are not the intended recipient, any dissemination, distribution or copying of this e-mail and any attachments is strictly prohibited. If you have received this e-mail in error, please immediately notify me and permanently delete the original and all copies and printouts of this e-mail and any attachments.


On Mar 17, 2019, at 11:24, jeffpinsky@... via Groups.Io <jeffpinsky@...> wrote:

Hi all,

?

I¡¯ve been half heartedly poking at getting my 85662 SA display unit (2112A02157) working on and off for years.? Has a dead display and blows the HV driver fuse.? Since I have 3 other SA¡¯s (don¡¯t ask why, it¡¯s a sickness), I really haven¡¯t given it a full blown effort.

?

It¡¯s spring cleaning time and I have to get rid of the projects I¡¯ll never get to.? I¡¯m parting out the display/IF unit.? It¡¯s too big and heavy to pack and ship.? I¡¯ve had some crappy experiences with swap and return on the auction site.

?

I have the boards below and offering them for $10 each (+ $10 for shipping on as many as you want).? Couldn¡¯t hurt to make an offer for a bit less.? If you buy 6, I¡¯ll thrown in whatever boards are left, so hurry. :) ? ?

?

A4A9 ? Control

A4A8 ? BW Filter

A4A7 ? 3Mhz Filter

A4A6A1/2 ? Up/Down Converter

A4A5? ? Step Gain

A4A4? ? BW Filter

A4A3? ? Log Amp/Filter

A4A2? ? Log Amp/Detector

A2A1? ? Video Processor

A3A9? ? Track & Hold

A3A8? ? A/D Converter

A3A7? ? Interface Assy

A3A6? ? Control Assy

A3A5? ? Data Manipulator

A3A4? ? Memory

A3A3? ? Line Generator

A3A2? ? Intensity Controller

A3A1? ? Trigger Assy

A1A4/5 & A1A2? ? Deflection (2) & CRT Controllers *

A1A7/A6/A8? Supply1/Rectifier/Supply2 *

A1A3 ? HV Supply + HV Multiplier? *

?

Any with more than one board on a line are as a group and count as one.? The last 3 groups with an asterisk (*) are where the problem lies.? The 23V supply fuse to the flyback driver blows. I never narrowed it down to exactly which board or component (may have even been the CRT) is the issue.? Since I couldn¡¯t test these with a dead display, they¡¯re as is.? Heck,? they¡¯re all as is; consider ¡®em for parts only.? I will say that in the course of troubleshooting, I did check the X/Y/Z outputs on the back and got a proper display on my scope when set to the test mode in the 85662.? That was before I misplaced the front panel (which I¡¯ve searched for and still can¡¯t find).

?

Let me know if you want any of the chassis carcass components, otherwise they¡¯ll get tossed.

?

Contact me here or off list via @ to wa2scl .

?

Thanks to reading.

?

Jeff


Re: 8662A output oddity

morestainless
 

Congratulations on your fine acquisition.? I just got mine going.? Got some voltage and calibration adjustments to do?


On Sun, Mar 17, 2019 at 3:32 PM, hpagilent@...
<hpagilent@...> wrote:
Hi, I recently acquired my first 8662A and while looking at its output on a spectrum analyzer I notice that there is a strange behavior that takes place in an approximate 2MHz span centered on 1024.8 MHz. At 1024.8 the noise floor increases by about 20dB and when tuning on either side of 1024.8 the noise floor decreases until it is completely back to normal at +/- 1 MHz on either side. I have included two pictures - one showing normal behavior and the other with the output tuned to 1024.8. This does not seem to happen at any other frequency in the unit's range. Has anyone else seen this? Thank you! 73, Jim N8ECI


8662A output oddity

 

Hi, I recently acquired my first 8662A and while looking at its output on a spectrum analyzer I notice that there is a strange behavior that takes place in an approximate 2MHz span centered on 1024.8 MHz. At 1024.8 the noise floor increases by about 20dB and when tuning on either side of 1024.8 the noise floor decreases until it is completely back to normal at +/- 1 MHz on either side. I have included two pictures - one showing normal behavior and the other with the output tuned to 1024.8. This does not seem to happen at any other frequency in the unit's range. Has anyone else seen this? Thank you! 73, Jim N8ECI


Re: Parting out hp85662A display unit

 

¿ªÔÆÌåÓý

Jeff,

Was the display for a ¡°B¡±? ?If so I would like all of the A4 cards please

Regards,

?

Stephen Hanselman

Datagate Systems, LLC

3107 North Deer Run Road #24

Carson City, Nevada, 89701

(775) 882-5117?office

(775) 720-6020?mobile

s.hanselman@...

a Service Disabled, Veteran Owned Small Business

DISCLAIMER:

This e-mail and any attachments are intended only for use by the addressee(s) named herein and may contain legally privileged and/or proprietary information. If you are not the intended recipient, any dissemination, distribution or copying of this e-mail and any attachments is strictly prohibited. If you have received this e-mail in error, please immediately notify me and permanently delete the original and all copies and printouts of this e-mail and any attachments.


On Mar 17, 2019, at 11:24, jeffpinsky@... via Groups.Io <jeffpinsky@...> wrote:

Hi all,

?

I¡¯ve been half heartedly poking at getting my 85662 SA display unit (2112A02157) working on and off for years.? Has a dead display and blows the HV driver fuse.? Since I have 3 other SA¡¯s (don¡¯t ask why, it¡¯s a sickness), I really haven¡¯t given it a full blown effort.

?

It¡¯s spring cleaning time and I have to get rid of the projects I¡¯ll never get to.? I¡¯m parting out the display/IF unit.? It¡¯s too big and heavy to pack and ship.? I¡¯ve had some crappy experiences with swap and return on the auction site.

?

I have the boards below and offering them for $10 each (+ $10 for shipping on as many as you want).? Couldn¡¯t hurt to make an offer for a bit less.? If you buy 6, I¡¯ll thrown in whatever boards are left, so hurry. :) ? ?

?

A4A9 ? Control

A4A8 ? BW Filter

A4A7 ? 3Mhz Filter

A4A6A1/2 ? Up/Down Converter

A4A5? ? Step Gain

A4A4? ? BW Filter

A4A3? ? Log Amp/Filter

A4A2? ? Log Amp/Detector

A2A1? ? Video Processor

A3A9? ? Track & Hold

A3A8? ? A/D Converter

A3A7? ? Interface Assy

A3A6? ? Control Assy

A3A5? ? Data Manipulator

A3A4? ? Memory

A3A3? ? Line Generator

A3A2? ? Intensity Controller

A3A1? ? Trigger Assy

A1A4/5 & A1A2? ? Deflection (2) & CRT Controllers *

A1A7/A6/A8? Supply1/Rectifier/Supply2 *

A1A3 ? HV Supply + HV Multiplier? *

?

Any with more than one board on a line are as a group and count as one.? The last 3 groups with an asterisk (*) are where the problem lies.? The 23V supply fuse to the flyback driver blows. I never narrowed it down to exactly which board or component (may have even been the CRT) is the issue.? Since I couldn¡¯t test these with a dead display, they¡¯re as is.? Heck,? they¡¯re all as is; consider ¡®em for parts only.? I will say that in the course of troubleshooting, I did check the X/Y/Z outputs on the back and got a proper display on my scope when set to the test mode in the 85662.? That was before I misplaced the front panel (which I¡¯ve searched for and still can¡¯t find).

?

Let me know if you want any of the chassis carcass components, otherwise they¡¯ll get tossed.

?

Contact me here or off list via @gmail.com to wa2scl .

?

Thanks to reading.

?

Jeff


Re: SG4 Service Manual

 

Hi All
The Wandel & Goltermann Museum offers visitors an insight into the career of the company Wandel & Goltermann from the founding of the company in 1923 until the 1990s.?The focus of the exhibition is on telecommunications measurement technology from the 1950s to the 1990s.?With the help of numerous exhibits and live applications, the development of telecommunications during this time is clearly conveyed to the visitor.
The purpose of the museum is to honor the entrepreneurial achievements of company founders Wolfram Wandel and Ulrich Goltermann and to keep them alive in the immediate environment - but also in the interested professional world.?The extensive company documentation gives an overview of the creation of Wandel & Goltermann from the foundation in 1923 until the merger with the American company Wavetek in 1998.
The former company with plants in Germany, the USA, England, France, Brazil and Switzerland as well as 25 own sales companies last employed 1,600 people worldwide and generated annual sales of around DM 500 million.?


Do a Google search for Wandel & Goltermann Museum.

George G6HIG



On Sunday, March 17, 2019 6:10 PM, nigel adams via Groups.Io <nigel.adams@...> wrote:


Hi,

The successor company to W&G is JDSU or Viavi...

Perhaps they have an archive somewhere, or like some modern companies don't care about their history and haven't saved anything.

Worth a try first though.


Sent from Samsung by banging a nail into a piece of wood.

-------- Original message --------
From: Alberto di Bene
Date:17/03/2019 13:05 (GMT+00:00)
Subject: [HP-Agilent-Keysight-equipment] SG4 Service Manual

Please excuse this my intrusion in an HP forum with a request non-HP-related, but I am in desperate need of the Service Manual of the Wandel & Goltermann SG4 Display Unit. Google is of no help...? I have a SG4 which needs some repairs. If somebody should have the PDF of that manual, please send it to me at the address dibene (at) usa.net

Many thanks in advance.

Alberto