Date

Re: HP 8600A partial manual needed

Brian Goldsmith
 

Greetings to all.I have a manual for the 8600A (LED display version) Digital
Marker but it lacks the schematics and PCB layouts.Can anyone help with a
scan or photocopy? All costs gratefully prepaid.
Brian Goldsmith.


Re: HP 3406A manual needed

Gary Beech
 

¿ªÔÆÌåÓý

Hi all I have a PDF of the HP3406a, it is 4MB can you accept a file this large
?
regards Gary Beech VK2KYP


From: aquaman8_2001 [mailto:aquaman8_2001@...]
Sent: Thursday, 11 November 2004 1:04 AM
To: hp_agilent_equipment@...
Subject: [hp_agilent_equipment] Re: HP 3406A manual needed


Hello,

If you do have that PDF, I would also really appreciate a copy!

Great group!!

Thanks in advance,

Mitch


--- In hp_agilent_equipment@..., "Gary Beech"
wrote:
> Hi Dave,
> I think I have a PDF of the HP3406a I will confirn later when I
get home
>?
> regards
>?
> Gary Beech VK2KYP
>
> ????? -----Original Message-----
> ????? From: daveolla [mailto:grobbins@n...]
> ????? Sent: Tue 9/11/2004 10:17 PM
> ????? To: hp_agilent_equipment@...
> ????? Cc:
> ????? Subject: [hp_agilent_equipment] HP 3406A manual needed
> ?????
> ?????
> ?????
> ????? Greetings, I am looking for a HP 3406A manual, that
hopefully is
> ????? cheaper than the actual unit was. PDF is plenty fine.
> ????? Thanks
> ????? Dave Robbins
> ?????
> ?????
> ?????
> ?????
> ?????
> Yahoo! Groups Sponsor?????
> ADVERTISEMENT
> click here
<
76/D=groups/S=1705083663:HM/EXP=1100085501/A=2376776/R=0/SIG=11ldm1jv
c/*> ?????
>? <
M=294855.5468653.6549235.3001176/D=groups/S=:HM/A=2376776/rand=591103
946> ?????
>
> ?????
>?? _____?
>
> ????? Yahoo! Groups Links
> ?????
>
> ????? *????? To visit your group on the web, go to:
> ????? ?????
> ????? ????? ?
> ????? *????? To unsubscribe from this group, send an email to:
> ????? ????? hp_agilent_equipment-unsubscribe@...
subject=Unsubscribe>
> ????? ????? ?
> ????? *????? Your use of Yahoo! Groups is subject to the Yahoo!
Terms of Service <> .
>
>
>
>
>
****
***
> The information contained in this e-mail message and any
accompanying files
> is or may be confidential.If you are not the intended recipient,
any use,
> dissemination, reliance,forwarding, printing or copying of this e-
mail or
> any attached files is unauthorised.This e-mail is subject to
copyright. No
> part of it should be reproduced,adapted or communicated without
the written
> consent of the copyright owner.If you have received this e-mail in
error,
> please advise the sender immediately by return e-mail, or
telephone and
> delete all copies.Fairfax does not guarantee the accuracy or
completeness
> of any information contained in this e-mail or attached files.
Internet
> communications are not secure, therefore Fairfax does not accept
legal
> responsibility for the contents of this message or attached files.
>
****
***






**
The information contained in this e-mail message and any accompanying files
is or may be confidential.If you are not the intended recipient, any use,
dissemination, reliance,forwarding, printing or copying of this e-mail or
any attached files is unauthorised.This e-mail is subject to copyright. No
part of it should be reproduced,adapted or communicated without the written
consent of the copyright owner.If you have received this e-mail in error,
please advise the sender immediately by return e-mail, or telephone and
delete all copies.Fairfax does not guarantee the accuracy or completeness
of any information contained in this e-mail or attached files. Internet
communications are not secure, therefore Fairfax does not accept legal
responsibility for the contents of this message or attached files.
**


Re: HP 3406A manual needed

 

Hello,

If you do have that PDF, I would also really appreciate a copy!

Great group!!

Thanks in advance,

Mitch

--- In hp_agilent_equipment@..., "Gary Beech"
<gbeech@m...> wrote:
Hi Dave,
I think I have a PDF of the HP3406a I will confirn later when I
get home

regards

Gary Beech VK2KYP

-----Original Message-----
From: daveolla [mailto:grobbins@n...]
Sent: Tue 9/11/2004 10:17 PM
To: hp_agilent_equipment@...
Cc:
Subject: [hp_agilent_equipment] HP 3406A manual needed



Greetings, I am looking for a HP 3406A manual, that
hopefully is
cheaper than the actual unit was. PDF is plenty fine.
Thanks
Dave Robbins





Yahoo! Groups Sponsor
ADVERTISEMENT
click here
<
76/D=groups/S=1705083663:HM/EXP=1100085501/A=2376776/R=0/SIG=11ldm1jv
c/*>
<
M=294855.5468653.6549235.3001176/D=groups/S=:HM/A=2376776/rand=591103
946>


_____

Yahoo! Groups Links


* To visit your group on the web, go to:


* To unsubscribe from this group, send an email to:
hp_agilent_equipment-unsubscribe@...
<mailto:hp_agilent_equipment-unsubscribe@...?
subject=Unsubscribe>

* Your use of Yahoo! Groups is subject to the Yahoo!
Terms of Service <> .




****
***
The information contained in this e-mail message and any
accompanying files
is or may be confidential.If you are not the intended recipient,
any use,
dissemination, reliance,forwarding, printing or copying of this e-
mail or
any attached files is unauthorised.This e-mail is subject to
copyright. No
part of it should be reproduced,adapted or communicated without
the written
consent of the copyright owner.If you have received this e-mail in
error,
please advise the sender immediately by return e-mail, or
telephone and
delete all copies.Fairfax does not guarantee the accuracy or
completeness
of any information contained in this e-mail or attached files.
Internet
communications are not secure, therefore Fairfax does not accept
legal
responsibility for the contents of this message or attached files.
****
***


Re: HP 3406A manual needed

Gary Beech
 

Hi Dave,
I think I have a PDF of the HP3406a I will confirn later when I get home

regards

Gary Beech VK2KYP

-----Original Message-----
From: daveolla [mailto:grobbins@...]
Sent: Tue 9/11/2004 10:17 PM
To: hp_agilent_equipment@...
Cc:
Subject: [hp_agilent_equipment] HP 3406A manual needed



Greetings, I am looking for a HP 3406A manual, that hopefully is
cheaper than the actual unit was. PDF is plenty fine.
Thanks
Dave Robbins





Yahoo! Groups Sponsor
ADVERTISEMENT
click here <>
<>


_____

Yahoo! Groups Links


* To visit your group on the web, go to:


* To unsubscribe from this group, send an email to:
hp_agilent_equipment-unsubscribe@... <mailto:hp_agilent_equipment-unsubscribe@...?subject=Unsubscribe>

* Your use of Yahoo! Groups is subject to the Yahoo! Terms of Service <> .




**
The information contained in this e-mail message and any accompanying files
is or may be confidential.If you are not the intended recipient, any use,
dissemination, reliance,forwarding, printing or copying of this e-mail or
any attached files is unauthorised.This e-mail is subject to copyright. No
part of it should be reproduced,adapted or communicated without the written
consent of the copyright owner.If you have received this e-mail in error,
please advise the sender immediately by return e-mail, or telephone and
delete all copies.Fairfax does not guarantee the accuracy or completeness
of any information contained in this e-mail or attached files. Internet
communications are not secure, therefore Fairfax does not accept legal
responsibility for the contents of this message or attached files.
**


Re: HP8558 / A8 sweep generator card schematic request

 

The manual on LOGSA includes the A8 board schematic (see page 8-53):


HP 3406A manual needed

 

Greetings, I am looking for a HP 3406A manual, that hopefully is
cheaper than the actual unit was. PDF is plenty fine.
Thanks
Dave Robbins


Manual for hp 5321B

 

I am looking for manual for hp counter 5321B. Any help would be appritiated.

Regards

Miroslav Pokorni


HP89440A Vector Signal Analyzer

Mark Myers
 

I am looking for information for programming NTIA and FCC Spectral masks into the HP89440A Vector Signal Analyzer.
Any help with this would be appreciated. Any information on a .sdf files would be good also. I have a file for one mask, if I can rename and edit that mask for a new mask that would be nice.

_________________________________________________________________
Get ready for school! Find articles, homework help and more in the Back to School Guide!


Re: Help with hp 141T and 8553

ke1ihcrash
 

John,

Thanks for the note. I will check them out, I did not even think to
check the supply voltages. I have an electronic version of an US
military manual and go look.

Thanks

John--- In hp_agilent_equipment@..., "John Miles"
<jmiles@p...> wrote:
That's pretty far off for an 8553. How are the power supply
voltages?

www.manualsplus.com will most likely have the book.

-- john KE5FX



Hi,

I recently picked up a used 141t with a 8552 if and a 8553 rf
front
end. the anlayzer works, but the center frequency is about 24
mhz
off. Does any one know a procedure recal the center frequency?
I
have tried to adjust the "tuning range" adjustment, that had no
effect.

Thanks

john



Re: Help with hp 141T and 8553

John Miles
 

That's pretty far off for an 8553. How are the power supply voltages?

www.manualsplus.com will most likely have the book.

-- john KE5FX


Hi,

I recently picked up a used 141t with a 8552 if and a 8553 rf front
end. the anlayzer works, but the center frequency is about 24 mhz
off. Does any one know a procedure recal the center frequency? I
have tried to adjust the "tuning range" adjustment, that had no
effect.

Thanks

john



Help with hp 141T and 8553

ke1ihcrash
 

Hi,

I recently picked up a used 141t with a 8552 if and a 8553 rf front
end. the anlayzer works, but the center frequency is about 24 mhz
off. Does any one know a procedure recal the center frequency? I
have tried to adjust the "tuning range" adjustment, that had no
effect.

Thanks

john


Re: 2430A & 2433 Power up Failures

 

Gearoid,

I believe if the 7000 CCD error comes up the CCD is faulty and the scope is a right off as the CCD 's are unobtainable. There are lots of these scopes laying around because there are no replacement parts.

David

At 01:42 AM 4/11/2004, you wrote:


Hi all,
I have a Tek 2432 which fails the following;
1: 7000 CCD
2: 8000 PA
3: 9000 TRIGS

I also have a Tek 2432A which fails;
1: 5000 WP
2: 8000 PA -- no pass/fail message
3: 9000 TRIGS -- no pass/fail message

Is this an almighty going round in circles job or has anyone been
through this before? Before you suggest it - the batteries are ok.
I may donate these if they require a lot of time to fix to somebody
who has the time.

Regards

Gearoid Loughnane
Dublin








Yahoo! Groups Links



David Hopkins (VK4ZF)
Queensland
AUSTRALIA
davhop@...
S 27, 22.294 E153, 11.008


Re: 2430A & 2433 Power up Failures

 

You might have better luck on a tektronix group.

-Chuck

gearoidloughnane wrote:

Hi all,
I have a Tek 2432 which fails the following;
1: 7000 CCD
2: 8000 PA
3: 9000 TRIGS
I also have a Tek 2432A which fails;
1: 5000 WP
2: 8000 PA -- no pass/fail message
3: 9000 TRIGS -- no pass/fail message
Is this an almighty going round in circles job or has anyone been through this before? Before you suggest it - the batteries are ok.
I may donate these if they require a lot of time to fix to somebody who has the time.
Regards
Gearoid Loughnane
Dublin
Yahoo! Groups Links


Re: 2430A & 2433 Power up Failures

frank Aras
 

It takes to much time to fix it and/or too much money. I am willing to take
them from you.
Regards
frank

-----Original Message-----
From: gearoidloughnane [mailto:echelon@...]
Sent: Wednesday, November 03, 2004 10:43 AM
To: hp_agilent_equipment@...
Subject: [hp_agilent_equipment] 2430A & 2433 Power up Failures



Hi all,
I have a Tek 2432 which fails the following;
1: 7000 CCD
2: 8000 PA
3: 9000 TRIGS

I also have a Tek 2432A which fails;
1: 5000 WP
2: 8000 PA -- no pass/fail message
3: 9000 TRIGS -- no pass/fail message

Is this an almighty going round in circles job or has anyone been
through this before? Before you suggest it - the batteries are ok.
I may donate these if they require a lot of time to fix to somebody
who has the time.

Regards

Gearoid Loughnane
Dublin








Yahoo! Groups Links


2430A & 2433 Power up Failures

gearoidloughnane
 

Hi all,
I have a Tek 2432 which fails the following;
1: 7000 CCD
2: 8000 PA
3: 9000 TRIGS

I also have a Tek 2432A which fails;
1: 5000 WP
2: 8000 PA -- no pass/fail message
3: 9000 TRIGS -- no pass/fail message

Is this an almighty going round in circles job or has anyone been
through this before? Before you suggest it - the batteries are ok.
I may donate these if they require a lot of time to fix to somebody
who has the time.

Regards

Gearoid Loughnane
Dublin


Re: Schematics for HP 54510A scope

Harlan Gustafson
 

¿ªÔÆÌåÓý

I'm giving this second hand so I'll find out more and get back.? There is supposed to be a 15v supply to the crt circuitry which is down.? Which schematics do we need?
Thanks,
Gus
?

----- Original Message -----
Sent: Monday, November 01, 2004 11:38 PM
Subject: [hp_agilent_equipment] Re: Schematics for HP 54510A scope


--- In hp_agilent_equipment@..., "Harlan Gustafson"
<accellodyne@n...> wrote:
> Does anyone know where I could find schematics for my HP 54510A
digital oscilloscope?? I've looked in a few places without luck.
> Thanks,
> Gus

could you be more specific about what it is you're looking for? are
you looking for board level diagrams or actual circuit schematics.
most of the older service/troubleshooting manuals at least had board
level troubleshooting diagrams. if you could give me some more info i
might be able to help you.





Re: Schematics for HP 54510A scope

 

Fred Olsen wrote:

NR1DX wrote:

HP and Tek both took the position that you had to buy a
new repayment cad ...
What's that?

Fred
That's where you have to take out a home loan to pay for the replacement card :)

John
<>


Re: Schematics for HP 54510A scope

Fred Olsen
 

new repayment cad ...
NR1DX wrote:
What that is a really bad typo
"replacement card"
That's OK, I understand. The hour was getting late, the bottle was getting low ... ;<))

Fred
--
<><
--
Outgoing checked by eTrust EZ AV


Re: Schematics for HP 54510A scope

NR1DX
 

Fred

What that is a really bad typo

Shuda been

"replacement card"

Dave

At 10:14 PM 11/1/2004, Fred Olsen wrote:

NR1DX wrote:
HP and Tek both took the position that you had to buy a
new repayment cad ...
What's that?

Fred
--
<><
--
Outgoing checked by eTrust EZ AV





Yahoo! Groups Links



Dave
NR1DX
nr1dx@...

"A man who picks a cat up by the tail learns a lesson he can learn no other way" .... Mark Twain


Re: Schematics for HP 54510A scope

 

--- In hp_agilent_equipment@..., "Harlan Gustafson"
<accellodyne@n...> wrote:
Does anyone know where I could find schematics for my HP 54510A
digital oscilloscope? I've looked in a few places without luck.
Thanks,
Gus
could you be more specific about what it is you're looking for? are
you looking for board level diagrams or actual circuit schematics.
most of the older service/troubleshooting manuals at least had board
level troubleshooting diagrams. if you could give me some more info i
might be able to help you.