Re: HP 6012B volt zero and current regulation
Hi Dave. I got to thinking a bit about the point Alan made below, and
toggle quoted message
Show quoted text
it's a very good one. If you consider the turn-on behavior of an unregulated linear supply (i.e. the classic transformer, bridge rectifier, filter capacitor construct), the shape of the voltage ramp-up curve should be dominated by the capacitor's charging characteristics. But if you place a regulator after that, the output voltage (of the regulator) will rise, and go potentially quite high above the setpoint until the regulation loop takes over. The width of the waveform above the setpoint voltage will correspond in inverse proportion to the bandwidth of the control loop of the regulator. Linear and switching regulators will both exhibit this behavior, though modern switching regulators with high switching frequencies and high-bandwidth control loops will keep it well under control. In practice this is usually so brief that it doesn't bother anything, a few dozen microseconds at most. I myself have watched it on an oscilloscope (I design and characterize a lot of voltage regulators) but have never seen it do any harm. -Dave On 08/20/2018 05:06 PM, David Speck wrote:
OK, I'll bite-- --
Dave McGuire, AK4HZ New Kensington, PA |
Re: HP 6012B volt zero and current regulation
¿ªÔÆÌåÓýOK, I'll bite-- What kind of power supply would you recommend?? I always thought
that a regulated supply would be the safest.? Were these linear or switcher supplies that produced the spikes? Thanks, Dave On 8/20/2018 2:31 PM, Alan Melia via
Groups.Io wrote:
|
Joel Kist, was Re: [HP-Agilent-Keysight-equipment] HP 6012B volt zero and current regulation
On 08/20/2018 02:48 PM, Mark Wendt wrote:
Off subject ¨C I still get duplicates ¨C Joel?Yes, he is. There have been no duplicate messages here. That's a Yahoo problem specifically. Joel, you are subscribed at both your sbcglobal.net address and your att.net address. If you let me know which one you want to keep, I can chop the other one for you. -Dave -- Dave McGuire, AK4HZ New Kensington, PA |
Re: A huge thank you!
kevin kearns
¿ªÔÆÌåÓýWell done David!
|
Re: HP 6012B volt zero and current regulation
¿ªÔÆÌåÓýI did a lot of power supply testing for HP, Agilent, and NGC, including the 6012A/B¡¯s,Here is the spec sheet:?. ?What you want to look at is the transient response - right hand column. ?Calibration transient response is tested for both voltage and current modes and is tested full to no resistive load, and also with 100/90% resistive load change. ?2ms/100mv. ?I believe that a min. 5 min. warmup time is required for specs to apply. Don Bitters |
Re: 8566B SA all PLLs unlocked
I too have seen a failed power resistor. ?Other failed item maybe be a leaky diode that then shuts down the power supply. ?If the -40v supply is other than 0, -40v suspect a leaky diode, or shorted or leaky cap on the affected supply. ?
You will also get a similar symptom if you have an ext 10 MHz ref and it drops out, or you lose the internal 10MHz, or the 100MHz loop is down. Don Bitters |
Re: HP 6012B volt zero and current regulation
On Mon, Aug 20, 2018, 14:22 Joel R Kist <joel-kist@...> wrote:
By chance are you subscribed to the group with two different email addresses? Check the headers of the emails that are dupes the next time you receive them.? Mark |
Re: HP 6012B volt zero and current regulation
?
toggle quoted message
Show quoted text
During 10 years or so doing life tests on
transistors in the 1960s and 70s I would not use regulated power supplies for
powering the devices during their life test. When asked, I demonstrated the
switch on pulse with a Tek storage scope. ok it was only microseconds wide but
peaked at the unregulated level.~35v .....enough to damage 24v transistors.This
can also happen with a transcient load. It is to do with the bandwidth of the
feedback loop.
?
Alan
G3NYK
?
|
Re: HP 6012B volt zero and current regulation
¿ªÔÆÌåÓýAlso many bench supplies have very large output capacitors. I have a smaller HP on my bench and recorded over 100 amps instantaneous on a short with a couple feet of #10 leads. That was all from the output cap, it¡¯s a 10 amp supply.?Peter On Aug 20, 2018, at 2:18 PM, george edmonds via Groups.Io <G6HIG@...> wrote:
|
Re: HP 6012B volt zero and current regulation
¿ªÔÆÌåÓýOff subject ¨C I still get duplicates ¨C Joel STS SS ? From: [email protected] [mailto:[email protected]] On Behalf Of wilson2115@...
Sent: Monday, August 20, 2018 12:50 PM To: [email protected] Subject: Re: [HP-Agilent-Keysight-equipment] HP 6012B volt zero and current regulation ? I refined my test... |
Re: HP 6012B volt zero and current regulation
Hi In my experience most bench power supplies show a similar behaviour. ?At one time I suffered repeated damage to equipment that was powered from a bench supply, when checked with a storage scope it went to a high voltage on turn on before settling back to the set voltage. I still think that for valid measurements you cannot use a light bulb, they are NOT a linear devices if you do an applied voltage plot against observed resistance, from a sixtey year old memory of an experiment they are square law devices 73 George G6HIG ?? On Monday, August 20, 2018 5:50 PM, "wilson2115@..." <wilson2115@...> wrote: I refined my test... Instead of ramping the voltage with the knob. I basically set the voltage to 20V and the amperage to 2 amps. I connected the voltmeter in series with the bulb and performed a dynamic load test by opening and closing the circuit by touching the probe lead of the voltmeter to the output of the power supply... I also have the Fluke voltmeter set to peak amperage measurement and these are the results that I found. With the 6012B set to 20V and 2A, upon closing the circuit the peak amps are 4 amps but quickly settles to 2 amps With the 6024A set to 20V and 2 A, upon closing the circuit the peaks amps are about 2.5 but quickly settles to 2 amps. The important observation here is the 6012B settles back to 2 amps, it seems the settle time is longer and the deviation is larger than the 6024A. Although these are two completely different supplies. If anything I could be nit picking here on the 6012B for its regulation settle time and overshoot, which is why I wanted to see if someone else had a similar supply and could see if they could duplicate the same results to see if this is normal... Thanks for the help |
Re: HP 6012B volt zero and current regulation
I refined my test...
Instead of ramping the voltage with the knob. I basically set the voltage to 20V and the amperage to 2 amps. I connected the voltmeter in series with the bulb and performed a dynamic load test by opening and closing the circuit by touching the probe lead of the voltmeter to the output of the power supply... I also have the Fluke voltmeter set to peak amperage measurement and these are the results that I found. With the 6012B set to 20V and 2A, upon closing the circuit the peak amps are 4 amps but quickly settles to 2 amps With the 6024A set to 20V and 2 A, upon closing the circuit the peaks amps are about 2.5 but quickly settles to 2 amps. The important observation here is the 6012B settles back to 2 amps, it seems the settle time is longer and the deviation is larger than the 6024A. Although these are two completely different supplies. If anything I could be nit picking here on the 6012B for its regulation settle time and overshoot, which is why I wanted to see if someone else had a similar supply and could see if they could duplicate the same results to see if this is normal... Thanks for the help |
Re: A huge thank you!
Thanks for the smooth move! On Mon, Aug 20, 2018 at 6:32 AM Keith Monahan via Groups.Io <keith=[email protected]> wrote: Thanks indeed! |