8594E Cal Signal not Found ?


 

After replacing battery I entered the Time Base Correction, Frequency Response Correction & Step Attenuator Corrections.
Now when I go to perform Cal Freq or Cal Amptd I’m getting a Cal Signal Not Found.
I checked the CAL OUT and I’m seeing 300.000824MHz -20dBm on my Spectrum Analyzer.
Input using an HP E4432B Signal Generator set 300MHz -20dBm I’m seeing 300.8MHz at -28dBm.
As I change the Span I get different frequency’s showing?
?
Andrew
?
?

Join [email protected] to automatically receive all group messages.