Re: 85719A Noise Figure Measurements Personality for HP8595A


 

On Thu, Mar 28, 2024 at 05:47 AM, vk2bea wrote:
HP8595A
Michael,?

You need the personality module reader that attaches to the rear of the 95A. The card is read by the?
reader and uses the noise marker mode and the noise measurement method is the GAIN method of?
noise figure techniques. Today, sometimes referred to as the cold source method. The SW optimizes
the SA to obtain a low noise capability, takes into account the NEBW, etc... You most likely will need?
an LNA in front of your analyzer particularly if you are trying to do low noise measurements. Required?
to avoid the 2nd stage NF contribution from the SA.?