**** * HP 8568B OPERATION VERIFICATION PROGRAM * **** GENERAL INFORMATION The following is a list of the tests performed by the Operation Verification and the equipment required for each test. TEST NO. TITLE EQUIPMENT REQUIRED ============================================ 1 INPUT ATTENUATOR SWITCHING UNCERTAINTY HP 3335A 2 AMPLITUDE FIDELITY (Log) HP 3335A 3 AMPLITUDE FIDELITY (Lin.) HP 3335A 4 LOG SCALE SWITCHING UNCERTAINTY HP 3335A 5 IF GAIN UNCERTAINTY HP 3335A 6 FREQUENCY SPAN ACCURACY HP 3335A 7 SWEEP TIME ACCURACY NONE 8 RESOLUTION BANDWIDTH NONE 9 RESIDUAL FM NONE 10 LINE RELATED SIDEBANDS NONE 11 RF GAIN UNCERTAINTY NONE 12 AVERAGE NOISE LEVEL HP 11593A 13 RESIDUAL RESPONSES HP 11593A 14 FREQUENCY RESPONSE HP 3335A HP 436A, Option 022 HP 8444A, Option 059 HP 8482A HP 11667A HP 8491B, Option 010 15 CAL OUTPUT AMPLITUDE ACCURACY HP 436A, Option 022 HP 8482A ============================================ The HP-IB addresses for the equipment are: HP 3335A Decimal Address 04 HP 436A, Option 022 Decimal Address 13 HP 8568B Decimal Address 18 If you wish to use addresses other than the ones listed, the instrument address assign- ments must be changed in the main program. The addresses are located near the top of the program and are clearly identified. The Operation Verification manual contains more important information concerning the veri- fication program. Some of the items covered in the manual are: * What the Operation Verfication program actually verifies. * What to do if you have trouble running the program. * What to do if the test instrument is out- of-tolerence. USING THE PROGRAM Getting Started After presentation of preliminary instruc- tions, the program advances to a pre-cali- bration routine. This routine is im- portant because the validity of the tests depend, in part, on proper instru- ment calibration. Please follow all instruc- tions as they appear on the spectrum analyzer CRT. Testing Single Tests Any of the tests may be performed singly by entering the test number followed by the spectrun analyzer data keyboard and push- ing Hz. The test is then performed before the program returns to the test menu. Repeat Tests Any of the test may be performed repeti- tively by entering the number of the test followed by kHz. The program continues to perform the specified test until the MHz key is pushed. This aborts te current test and returns the program to test menu. All Tests All fifteen tests may be performed in sequence by entering 0 Hz on the data key- board. The fifteen tests are performed in the order listed. Whenever an equipment change is necessary, a new test setup is displayed on the spectrum analyzer CRT and the program is halted until new con- nections are completed. Test Results The test number, title, specification, and measured data are printed on an HP-IB printer or on the spectrum analyzer CRT depending on your selection. If no printout is selected, only the test menu is updated. If the measured data indicates that a para- meter is out-of-tolerence, a double astrisk (**) is placed adjacent to that data. The printout also lists the title of the Per- formance Test (in the Tests and Adjustments Manual) related to the parameter that is out-of-tolerance. Verification of an instrument failure requires that this Performance Test be performed. Other Instructions All necessary instructions for proper operation of this program are displayed on the spectrum analyzer CRT as they are needed. Follow all instructions as they appear and remember that the current test can be aborted by pushing MHz.